Aubert A E, Vrolix M, De Geest H, Van de Werf F
Department of Cardiology, University Hospital Gasthuisberg, Catholic University Leuven, Belgium.
Int J Clin Monit Comput. 1995 May;12(2):77-83. doi: 10.1007/BF01142487.
Experimental findings are presented of an in vivo comparison between a Sentron catheter and another tip transducer manometer: a Millar microtip catheter. Both catheters have been used simultaneously in the left ventricle of dogs. Pressure variations were elicited by drug infusion. Pressure values and derivatives obtained from both systems were compared. A cross correlation between episodes of the two pressures was computed. Results from this study showed good correlation between left ventricular systolic pressure measured with both manometers (R = 0.992, p < 0.0001), end-diastolic pressure (R = 0.809, p < 0.0001) and between first derivatives: positive derivative (R = 0.993, p < 0.0001) and negative (R = 0.634, p < 0.0001). The mean cross correlation between both pressure signals was 0.61 +/- 0.04. In the frequency domain no statistical difference was found between the location of the maxima of the peaks. It is concluded that a Sentron manometer can be a valid alternative, at a reasonable price, to a cheaper, though less accurate fluid filled catheter and a more expensive 'golden standard' microtip catheter.
本文展示了Sentron导管与另一种尖端换能器压力计(Millar微尖端导管)之间的体内比较实验结果。两种导管同时用于犬的左心室。通过药物输注引发压力变化。比较了两个系统获得的压力值和导数。计算了两种压力事件之间的互相关。本研究结果表明,两种压力计测量的左心室收缩压之间具有良好的相关性(R = 0.992,p < 0.0001),舒张末期压力之间(R = 0.809,p < 0.0001)以及一阶导数之间:正导数(R = 0.993,p < 0.0001)和负导数(R = 0.634,p < 0.0001)。两种压力信号之间的平均互相关为0.61 +/- 0.04。在频域中,峰值最大值的位置之间未发现统计学差异。结论是,Sentron压力计可以以合理的价格成为一种有效的替代方案,替代价格较低但准确性较差的充液导管和更昂贵的“金标准”微尖端导管。