Shoucri R M, Pouliot M
Clin Chem. 1977 Sep;23(9):1527-30.
The Jaffé reaction for creatinine assay appears to follow pseudo-first-order kinetics; first-order rate constants are different for different samples. Rate constants for 10 different serum samples varied from a low value of 0.0040 +/- 0.0003 s-1 to 0.0084 +/- 0.0008 s-1. We describe an approach for determining first-order rate constants from kinetic data and discuss the effects of the above observations on the mathematical formulations required for reliable kinetic determinations of creatinine.
用于肌酐测定的杰氏反应似乎遵循准一级动力学;不同样品的一级速率常数不同。10个不同血清样品的速率常数范围从低值0.0040±0.0003 s⁻¹到0.0084±0.0008 s⁻¹。我们描述了一种从动力学数据确定一级速率常数的方法,并讨论了上述观察结果对可靠进行肌酐动力学测定所需数学公式的影响。