Importance of calibration for diameter and area determination by intravascular ultrasound.

作者信息

Stähr P, Rupprecht H J, Voigtländer T, Kearney P, Erbel R, Koch L, Krass S, Brennecke R, Meyer J

机构信息

2nd Medical Clinic, Johannes-Gutenberg-University, Mainz, Germany.

出版信息

Int J Card Imaging. 1996 Dec;12(4):221-9. doi: 10.1007/BF01797734.

Abstract

BACKGROUND

Intravascular ultrasound (IVUS) permits quantitative assessment of the lumen diameter and area of coronary arteries. The experimental study was performed to evaluate the accuracy of diameter and area measurements.

METHODS AND RESULTS

Lumen quantitation (lumen diameter D and cross-sectional area A) in lucite tubes (lumen diameter 2.5 to 5.7 mm, Plexiglas) was performed using a mechanical IVUS system (HP console, 3.5 F catheter, Boston Scientific, 30 MHz). The influence of fluid type (blood, water and saline solution), fluid temperature (20 degrees C/37 degrees C), catheter to catheter variation, gain setting and ultrasound frequency (12, 20 and 30 MHz) was determined. In blood at 20 degrees C there was a constant deviation of the measured diameter from the true luminal diameter of -0.29 +/- -0.04 mm (p < 0.06). In water and saline solution at 20 degrees C the mean deviation from true diameter was -0.21 +/- -0.06 mm (p < 0.06). At 37 degrees C, the deviation in blood was greater than at 20 degrees (-0.34 +/- -0.02 mm) which is > 10% in a 3 mm tube (p < 0.06). Three of the ten catheters tested in water at 20 degrees C underestimated true diameter by more than -0.3 mm. The deviation from true diameter (5 mm tube) with varying gain settings was -0.14 mm to -0.23 mm compared to -0.19 mm at standard settings (p > 0.288). At 12 MHz diameter measured was over-estimated. The error in absolute area estimation increased with increasing diameter tested in blood at 37 degrees C (-1.21 to -2.72 mm2), whereas the relative error ([Measured Area-True Area]/True Area x 100 [%]) was more striking at smaller diameters (up to -25% in the 2.5 mm tube).

CONCLUSION

Luminal diameters and areas are underestimated by this particular IVUS system. When IVUS imaging and measurements are made during coronary interventions this error should be taken into account with regard to appropriate sizing of the device and the assessment of the postprocedure result. Because systematic errors might also occur in other IVUS system (not tested in this study), it is advisable to ensure that each system is validated prior to clinical use, especially when exact measurements are required.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验