Pacing Clin Electrophysiol. 1997 Feb;20(2 Pt 1):359-63. doi: 10.1111/j.1540-8159.1997.tb06183.x.
The aim of the study was to observe changes over time of the stimulation requirements in spinal cord stimulation (SCS). Of 60 patients treated with SCS, 25 patients were selected because they had neurostimulators capable of measuring impedance noninvasively, and had not experienced electrode displacement. All 25 patients had Medtronic Pisces Quad 3487A (Medtronic, Inc.) neuroelectrodes with the tip positioned in the thoracic epidural space. The accuracy of the neurostimulators impedance measuring circuit was investigated in a test circuit. The error was < 11%. Stimulus requirements and impedance were recorded at the implantation and at follow-ups during a period of 24 months. During the first month after implantation, the stimulus requirements for satisfactory effect varied between +406% and -34%. After that period, only minor deviations were observed in most patients. To optimize the pain reducing effect of the spinal cord stimulation, frequent follow-ups are recommended during the first month; later on, the follow-up intervals can be extended. No tolerance development or pain resistance developed during SCS treatment.
本研究的目的是观察脊髓刺激(SCS)中刺激需求随时间的变化。在接受SCS治疗的60例患者中,选择了25例患者,因为他们拥有能够无创测量阻抗且未发生电极移位的神经刺激器。所有25例患者均使用美敦力双鱼座四极3487A(美敦力公司)神经电极,电极尖端置于胸段硬膜外间隙。在测试电路中研究了神经刺激器阻抗测量电路的准确性。误差<11%。在24个月的时间内,在植入时和随访时记录刺激需求和阻抗。植入后的第一个月,获得满意效果所需的刺激变化在+406%至-34%之间。在此之后,大多数患者仅观察到微小偏差。为了优化脊髓刺激的止痛效果,建议在第一个月进行频繁随访;之后,随访间隔可以延长。在SCS治疗期间未出现耐受性发展或疼痛抵抗。