J Lab Clin Med. 1977 Nov;90(5):856-62.
A simple gas chromatographic technique for analyzing H2 in expired air with a compact, relatively inexpensive gas chromatography specifically adapted for H2 analysis is described. Modifications in the basic chromatograph included increasing the capacity of the sample loop, lengthening the column, and using argon as the carrier gas to increase the sensitivity of the system. Methods for collecting and storing respiratory gas samples and for quantifying the results are described for clinical studies of carbohydrate malabsorption. Small errors in collection and quantitation are introduced by the method, but they are minor in relation to the changes in H2 concentration which are seen with significant malabsorption. This simplification of chromatograph technology could increase the general availability of this convenient, noninvasive and well-tolerated test of intestinal absorption of carbohydrates such as lactose.
描述了一种简单的气相色谱技术,该技术使用一种紧凑、相对便宜且专门适用于氢气分析的气相色谱仪来分析呼出气体中的氢气。对基本色谱仪的改进包括增加样品环的容量、延长色谱柱以及使用氩气作为载气以提高系统的灵敏度。文中还介绍了用于碳水化合物吸收不良临床研究的呼吸气体样本收集和储存方法以及结果定量方法。该方法在收集和定量过程中会引入小误差,但与明显吸收不良时氢气浓度的变化相比,这些误差较小。这种色谱技术的简化可以提高这种方便、无创且耐受性良好的乳糖等碳水化合物肠道吸收测试的普遍可用性。