Chronister C L
Eye Institute of the Pennsylvania College of Optometry, Philadelphia, USA.
Optom Vis Sci. 1997 Mar;74(3):164-6. doi: 10.1097/00006324-199703000-00028.
Although Schiotz tonometry has been long ago replaced by Goldmann applanation tonometry measurement of intraocular pressure, it remains a viable clinical technique in nursing homes and other "out of office" settings. Because Schiotz tonometers come into contact with the cornea and tear film, they must be properly disinfected to prevent cross-infection with HIV and other infectious pathogens. The Centers for Disease Control and Prevention (CDC) recommends that all eye contact devices be disinfected by a 5- to 10-min soak in one of following disinfectants: 3% hydrogen peroxide, 1:10 sodium hypochlorite (household bleach), or 70% isopropyl alcohol. The effects of these disinfectants on Schiotz tonometers have not been previously investigated. This study investigated the effects of CDC-recommended disinfectants on the structural integrity of Schiotz tonometers. The results of the study indicated that soaking in 1:10 sodium hypochlorite and 3% hydrogen peroxide damaged Schiotz tonometers and rendered them inoperable. Alcohol caused minor damage to the tonometer and was the safest of the CDC disinfectants. This study showed there is limited application of CDC infection control criteria to Schiotz tonometers due to the extensive corrosion and damage caused by hydrogen peroxide and bleach.
尽管眼压测量早已被戈德曼压平眼压测量法所取代,但在疗养院和其他“非医疗机构”环境中,它仍是一种可行的临床技术。由于眼压计会接触角膜和泪膜,因此必须进行适当消毒,以防止艾滋病毒和其他传染性病原体的交叉感染。疾病控制与预防中心(CDC)建议,所有眼部接触设备应浸泡在以下消毒剂之一中5至10分钟进行消毒:3%过氧化氢、1:10次氯酸钠(家用漂白剂)或70%异丙醇。这些消毒剂对眼压计的影响此前尚未进行过研究。本研究调查了CDC推荐的消毒剂对眼压计结构完整性的影响。研究结果表明,浸泡在1:10次氯酸钠和3%过氧化氢中会损坏眼压计并使其无法使用。酒精对眼压计造成的损害较小,是CDC消毒剂中最安全的。这项研究表明,由于过氧化氢和漂白剂会造成广泛的腐蚀和损坏,CDC感染控制标准在眼压计上的应用有限。