Sood D, Honavar S G
L.V. Prasad Eye Institute, Hyderabad, India.
Indian J Ophthalmol. 1998 Jun;46(2):113-6.
Precautions to prevent spread of infection through tonometers and gonioscopes are described in this article. Tonometers and gonioscopes should not be used in the presence of clinically manifest conjunctivitis and keratitis. The Schiotz tonometer should be dipped in a 1:1000 merthiolate solution, and rinsed in saline/distilled water prior to use. The Goldmann applanation prism tip can be wiped with gauze soaked in 70% isopropyl alcohol and then dried before use. Gonioscopes should be cleaned in running water, wiped with gauze soaked in 70% isopropyl alcohol, and then dried before use. Koeppes and goniotomy lenses can be sterilized with ethylene oxide, prior to use in surgery.
本文介绍了预防眼压计和前房角镜传播感染的注意事项。在存在临床表现的结膜炎和角膜炎时,不应使用眼压计和前房角镜。在使用前,施-氏眼压计应浸入1:1000硫柳汞溶液中,并用生理盐水/蒸馏水冲洗。戈德曼压平棱镜尖端可用浸有70%异丙醇的纱布擦拭,然后在使用前干燥。前房角镜应用流水清洗,用浸有70%异丙醇的纱布擦拭,然后在使用前干燥。在手术中使用前,克佩氏镜和前房角切开镜可用环氧乙烷消毒。