Estimation of contact tympanic membrane temperature with a noncontact infrared thermometer.

作者信息

Terndrup T E, Crofton D J, Mortelliti A J, Kelley R, Rajk J

机构信息

Department of Emergency Medicine, State University of New York Health Science Center at Syracuse, 13210, USA.

出版信息

Ann Emerg Med. 1997 Aug;30(2):171-5. doi: 10.1016/s0196-0644(97)70138-0.

Abstract

STUDY OBJECTIVE

To investigate the agreement between contact tympanic membrane (TM) and noncontact infrared ear thermometers in children.

METHODS

Twenty-three children (ages .5 to 10 years) undergoing elective tympanostomy tube placement were studied. An assistant used standard technique to record temperature with an infrared ear thermometer before and after TM temperature was obtained with a bead thermistor placed against the anterior-inferior quadrant of the TM.

RESULTS

Mean temperatures were not significantly different: initial IR ear, 36.66 degrees +/-.33 degrees C; TM, 36.71 degrees +/- .42 degrees C; final infrared ear, 36.57 degrees +/- .33 degrees C. The mean bias (difference between initial individual IR ear and TM temperatures) of -.05 degrees +/-.29 degrees C and the 95% limits of agreement of +.53 degrees to -.63 degrees C indicate an acceptable confidence (error range within 3.2% of average TM temperature) for use of the initial infrared ear temperature as an estimate of TM temperature.

CONCLUSION

The IR ear thermometer provides an accurate estimate of TM temperature in healthy children and may accurately reflect core body temperature.