A novel magnetic resonance imaging technique (STAR-HASTE) based on pulsed arterial spin labeling using a single shot acquisition method is described for perfusion imaging. The method is similar to EPISTAR in using STAR (Signal Targeting with Alternating Radiofrequency) technique for pulsed radiofrequency labeling of inflowing blood, but uses a half-Fourier single shot turbo spin-echo (HASTE) sequence for data acquisition instead of echo-planar imaging (EPI). Our preliminary studies show that STAR-HASTE permits perfusion imaging to be performed without many of the artifacts encountered with other imaging methods based on EPI acquisition. The novel method not only provides similar perfusion information to that obtained by EPISTAR, as demonstrated in the functional brain imaging study, but also eliminates magnetic susceptibility artifacts and image distortion commonly observed in EPI images. Furthermore, this technique can be readily implemented in MR systems without EPI capability.
本文描述了一种基于脉冲动脉自旋标记的新型磁共振成像技术(STAR-HASTE),该技术采用单次采集方法进行灌注成像。该方法与EPISTAR类似,使用STAR(交替射频信号靶向)技术对流入血液进行脉冲射频标记,但使用半傅里叶单次激发快速自旋回波(HASTE)序列进行数据采集,而非回波平面成像(EPI)。我们的初步研究表明,STAR-HASTE能够在不产生许多基于EPI采集的其他成像方法所遇到的伪影的情况下进行灌注成像。这项新技术不仅如在脑功能成像研究中所证明的那样,能提供与EPISTAR相似的灌注信息,还消除了EPI图像中常见的磁敏感性伪影和图像失真。此外,该技术可在不具备EPI功能的磁共振系统中轻松实现。