Lelièvre H, Gimenez M, Vandenesch F, Reinhardt A, Lenhardt D, Just H M, Pau M, Ausina V, Etienne J
Laboratoire de Bactériologie, Hôpital Louis Pradel, Lyon, France.
Eur J Clin Microbiol Infect Dis. 1997 Sep;16(9):669-74. doi: 10.1007/BF01708557.
In a study comparing the Bactec 9240 (Plus Aerobic/F and Anaerobic/F bottles, containing resins; Becton Dickinson, USA) and the Vital (standard aerobic and anaerobic bottles, with no additives; bioMérieux, France) blood culture systems, 6456 sets of four bottles of 9660 blood cultures submitted were evaluated. There were 531 clinically significant isolates from 795 positive blood cultures. Of the 531 positive blood cultures, 355 were positive in both systems, 141 with the Bactec 9240 alone, and 30 with the Vital alone (p < 0.001); five were not detected by either system. The average time to detection of positive cultures for the matched sets was 10.65 h and 18.41 h by the Bactec 9240 system and the Vital system, respectively. The false-positive rate per bottle was 0.65% in the Bactec 9240 and 0.71% in the Vital. The rate of false-negative pairs (i.e., major errors) was very low (0.12% for the Bactec 9240, 0.19% for the Vital) and not significantly different between the two systems. The striking differences in recovery of microorganisms may be due to the presence of resins in the Bactec medium. However, the observed superiority of the Bactec 9240, even for patients not receiving antibiotics, suggests that resins adsorb other inhibitors present in patients' blood.
在一项比较Bactec 9240(含树脂的需氧/F和厌氧/F瓶;美国BD公司)和Vital(无添加剂的标准需氧和厌氧瓶;法国生物梅里埃公司)血培养系统的研究中,对提交的6456套共四瓶9660份血培养样本进行了评估。在795份阳性血培养样本中分离出531株具有临床意义的菌株。在这531份阳性血培养样本中,两种系统均呈阳性的有355份,仅Bactec 9240呈阳性的有141份,仅Vital呈阳性的有30份(p<0.001);两种系统均未检测到5份样本。匹配样本集阳性培养物的平均检测时间,Bactec 9240系统为10.65小时,Vital系统为18.41小时。Bactec 9240每瓶的假阳性率为0.65%,Vital为0.71%。假阴性对(即主要错误)的发生率非常低(Bactec 9240为0.12%,Vital为0.19%),两种系统之间无显著差异。微生物回收率的显著差异可能是由于Bactec培养基中存在树脂。然而,即使对于未接受抗生素治疗的患者,Bactec 9240所观察到的优势表明树脂吸附了患者血液中存在的其他抑制剂。