Middleton A M, Chadwick M V, Gaya H
Department of Microbiology, Royal Brompton Hospital, London.
J Hosp Infect. 1997 Oct;37(2):137-43. doi: 10.1016/s0195-6701(97)90183-3.
The efficacy of 0.35% stabilized buffered peracetic acid solution ('Nu-Cidex') against clinical isolates of Mycobacterium tuberculosis, Mycobacterium avium-intracellulare and Mycobacterium chelonae in homogenized sputum was tested. An in-use method, using an automated bronchoscope washing machine, showed that over 10 cycles, at a disinfectant contact time of 5 min, M. tuberculosis and M. chelonae were effectively eradicated from the bronchoscope, even in the absence of detergent and pre-cleaning. M. avium-intracellulare was not eradicated in only one of 10 cycles with contact times of 5 and 10 min, but numbers were reduced by > 7 log10 and > 5 log10, respectively. With detergent present, M. avium-intracellulare was successfully eradicated in all cycles at a contact time of 5 min or greater. The results demonstrate that peracetic acid is an effective mycobactericidal agent for use in the disinfection of bronchoscopes.