Sijbers J, den Dekker A J, Van Audekerke J, Verhoye M, Van Dyck D
Department of Physics, University of Antwerp, Belgium.
Magn Reson Imaging. 1998;16(1):87-90. doi: 10.1016/s0730-725x(97)00199-9.
Magnitude magnetic resonance data are Rician distributed. In this note a new method is proposed to estimate the image noise variance for this type of data distribution. The method is based on a double image acquisition, thereby exploiting the knowledge of the Rice distribution moments.
幅度磁共振数据呈莱斯分布。在本笔记中,提出了一种新方法来估计此类数据分布的图像噪声方差。该方法基于双图像采集,从而利用莱斯分布矩的知识。