Suppr超能文献

评估HemoCue血糖仪用于新生儿低血糖即时诊断的效果。

Evaluation of HemoCue Blood Glucose Analyzer for the instant diagnosis of hypoglycaemia in newborns.

作者信息

Dahlberg M, Whitelaw A

机构信息

Department of Clinical Chemistry and Neurochemistry, SU/Mölndal, Sweden.

出版信息

Scand J Clin Lab Invest. 1997 Dec;57(8):719-24. doi: 10.3109/00365519709105234.

Abstract

The aim of the study was to evaluate a portable photometer, HemoCue Blood Glucose Analyzer, in the instant diagnosis of hypoglycaemia in newborns. The HemoCue is a simple, easy-to-handle photometer; with an analysis time of less than 240 s, it utilizes a modified glucose dehydrogenase method in 5 microliters whole blood. The HemoCue method was compared to a hexokinase method for deproteinized whole blood in a total of 118 samples from 58 newborns. The linear regression for these samples was Y = 1.19 x -1.02 (range 0.7-7.2 mmol/L), r = 0.90. Ten samples were < or = 2.0 mmol/L with both methods and 37 samples were < or = 2.0 mmol/L with the HemoCue method. The average difference (D) for each sample (n = 118) and the standard deviation (SD) for the difference were 0.45 +/- 0.46 mmol/L. Blood samples with a mean value with both methods < or = 2.0 mmol/L (n = 20) had a D and SD of 0.71 +/- 0.29 mmol/L. When testing for linearity at low glucose concentrations, the HemoCue method gave significantly lower values compared to an ideal line. The HemoCue method has several advantages in the analysis of glucose in newborns: short analysis time, small sample size, and no influence from glycolysis. However, in our investigation, falsely low values occurred, especially in the low measuring range, so the HemoCue method is not suitable in the diagnosis of hypoglycaemia in newborns.

摘要

本研究的目的是评估一款便携式光度计——HemoCue血糖仪,用于新生儿低血糖的即时诊断。HemoCue是一款简单、易于操作的光度计;分析时间少于240秒,它在5微升全血中采用改良的葡萄糖脱氢酶法。将HemoCue法与己糖激酶法用于58例新生儿的118份去蛋白全血样本进行比较。这些样本的线性回归方程为Y = 1.19x - 1.02(范围0.7 - 7.2 mmol/L),r = 0.90。两种方法检测时,有10份样本血糖值≤2.0 mmol/L,HemoCue法检测时有37份样本血糖值≤2.0 mmol/L。每个样本(n = 118)的平均差值(D)及差值的标准差(SD)为0.45±0.46 mmol/L。两种方法均值均≤2.0 mmol/L的血样(n = 20),其D和SD为0.71±0.29 mmol/L。在低葡萄糖浓度下检测线性时,与理想曲线相比,HemoCue法得出的值显著更低。HemoCue法在新生儿血糖分析方面有几个优点:分析时间短、样本量小且不受糖酵解影响。然而,在我们的研究中,出现了假性低值,尤其是在低测量范围内,所以HemoCue法不适用于新生儿低血糖的诊断。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验