Measurement of intraocular pressure with a modern noncontact tonometer.

作者信息

Koçak I, Orgül S, Saruhan A, Haefliger I, Hendrickson P, Flammer J

机构信息

University Eye Clinic, Basel, Switzerland.

出版信息

Ophthalmologica. 1998;212(2):81-7. doi: 10.1159/000027284.

Abstract

PURPOSE

To evaluate the intraocular pressure (IOP) in a population without known glaucoma as measured by means of a noncontact tonometer (NCT).

METHODS

IOP values of 842 persons, measured by means of a Pulsair 2000 NCT (Keeler Instruments, Inc., Broomhall, Pennsylvania) by 5 investigators were evaluated. Three consecutive measurements were taken first on the right eye and then on the left eye in each subject. Topical anesthesia was not used.

RESULTS

The distribution of the measured IOP values was significantly different from a normal distribution. The median IOP was 15.0 mmHg (lower quartile/ upper quartile 12.5/17.2). There were no differences in IOP values between male and female subjects or between right and left eyes. There was no correlation between age and IOP values. Ninety percent of the subjects had an IOP within a range of 9-24 mmHg in both eyes. Among the examined subjects, the IOP was above 24 and 21 mmHg in 4,6 and 11%, respectively, in at least one eye. The median range of consecutive IOP measurements in one eye was 3 mmHg (lower quartile/upper quartile 2.0/4.0). The median interocular IOP difference was 1.7 mmHg (0.7/2.7). The median individual coefficient of variation (CV) values were 10.0% for the right eyes, -10.8% for the left eyes. Among the 5 different examiners, the IOP values, range of 3 consecutive IOP readings in one eye, interocular asymmetry and CV values were statistically comparable.

CONCLUSION

The Pulsair 2000 NCT is very simple to use and its handling is mostly independent of the operator. The results of the present investigation might be of some value for future screening programs.

摘要

目的

通过非接触眼压计(NCT)评估无青光眼病史人群的眼压(IOP)。

方法

对5名研究人员使用Pulsair 2000 NCT(Keeler Instruments, Inc., Broomhall, Pennsylvania)测量的842人的眼压值进行评估。每位受试者先右眼后左眼连续测量3次。未使用表面麻醉。

结果

所测眼压值的分布与正态分布显著不同。眼压中位数为15.0 mmHg(下四分位数/上四分位数为12.5/17.2)。男性和女性受试者之间以及右眼和左眼之间的眼压值无差异。年龄与眼压值之间无相关性。90%的受试者双眼眼压在9 - 24 mmHg范围内。在受检者中,至少一只眼的眼压分别有4%、6%和11%高于24 mmHg和21 mmHg。单眼连续眼压测量的中位数范围为3 mmHg(下四分位数/上四分位数为2.0/4.0)。双眼眼压差异的中位数为1.7 mmHg(0.7/2.7)。右眼的个体变异系数(CV)中位数为10.0%,左眼为 - 10.8%。在5名不同的检查者中,眼压值、单眼连续3次眼压读数的范围、双眼不对称性和CV值在统计学上具有可比性。

结论

Pulsair 2000 NCT使用非常简单,其操作大多独立于操作人员。本研究结果可能对未来的筛查项目有一定价值。