Rohlmann A, Bergmann G, Graichen F, Mayer H M
Department of Biomechanics, Free University of Berlin, Germany.
Spine (Phila Pa 1976). 1998 Mar 1;23(5):537-42. doi: 10.1097/00007632-199803010-00005.
The loads acting on an internal spinal fixation device were measured in vivo.
To determine the influence of muscle forces on implant loads.
Only limited information exists regarding the loads acting on spinal implants in vivo. Though the muscles greatly influence spinal load, they have been neglected in most studies.
Telemeterized internal spinal fixation devices were used to study the influence of muscle forces on the implant loads in three patients before and after anterior interbody fusion.
Contracting abdominal or back muscles in a lying position was found to significantly increase implant loads. Hanging by the hands from wall bars as well as balancing with the hands on parallel bars reduced the implant loads compared with standing; however, hanging by the feet with the head upside down did not reduce implant loads compared with lying in a supine position. When lying on an operating table with only the foot end lowered so that the hips were bent, the patient had different load measurements in the conscious and anesthetized state before anterior interbody fusion. The anesthetized patient evidenced predominately extension moments in both fixators, whereas flexion moments were observed in the right fixator of the conscious patient. After anterior interbody fusion had occurred, the differences in implant loads resulting from anesthesia were small.
The muscles greatly influence implant loads. They prevent an axial tensile load on the spine when part of the body weight is pulling, e.g., when the patient is hanging by his hands or feet. The implant loads may be strongly altered when the patient is under anesthesia.
在体内测量作用于脊柱内固定装置的负荷。
确定肌肉力量对植入物负荷的影响。
关于体内作用于脊柱植入物的负荷,现有信息有限。尽管肌肉对脊柱负荷有很大影响,但在大多数研究中它们被忽视了。
使用遥测式脊柱内固定装置研究三名患者在前后路椎间融合术前和术后肌肉力量对植入物负荷的影响。
发现仰卧位时收缩腹部或背部肌肉会显著增加植入物负荷。与站立相比,双手抓住墙杆悬挂以及双手在双杠上保持平衡会降低植入物负荷;然而,与仰卧位相比,头朝下倒挂双脚并没有降低植入物负荷。当仅降低手术台的脚端使臀部弯曲时,患者在前后路椎间融合术前的清醒和麻醉状态下有不同的负荷测量结果。麻醉患者在两个固定器中主要表现为伸展力矩,而清醒患者的右固定器中观察到屈曲力矩。前路椎间融合术后,麻醉引起的植入物负荷差异很小。
肌肉对植入物负荷有很大影响。当部分体重牵拉时,例如患者用手或脚悬挂时,它们可防止脊柱受到轴向拉伸负荷。患者在麻醉状态下,植入物负荷可能会发生很大变化。