A SEM study of in vivo accuracy of the Root ZX electronic apex locator.

作者信息

Pagavino G, Pace R, Baccetti T

机构信息

Department of Endodontics, University of Florence, Italy.

出版信息

J Endod. 1998 Jun;24(6):438-41. doi: 10.1016/S0099-2399(98)80029-0.

Abstract

Twenty-nine unprepared, vital teeth were used to evaluate the accuracy of the Root ZX in locating the root canal foramen. After extraction of the teeth, a scanning electron microscope analyzed the relation of the file tip to the foramen. The sample was divided into two groups (Group A and Group B) according to the presence of a normal apical foramen (along the root main axis) or of a lateral foramen (deviating from the root main axis). With a tolerance level of +/- 0.5 mm, a clinical accuracy rate of 82.75% was recorded in the total sample. With a +/- 1.0-mm tolerance level, an accuracy of 100% was found. The error in locating the apex was significantly smaller in cases with a normal apical foramen (Group A) than in cases with a lateral foramen (Group B) (p < 0.001). An accuracy of 100% at the +/- 0.5-mm tolerance level was registered in Group A.

摘要

使用29颗未经处理的活髓牙来评估Root ZX定位根管口的准确性。牙齿拔除后,用扫描电子显微镜分析锉尖与根管口的关系。根据根尖孔(沿牙根主轴)或侧支根管(偏离牙根主轴)的情况将样本分为两组(A组和B组)。在±0.5毫米的公差水平下,总样本的临床准确率为82.75%。在±1.0毫米的公差水平下,准确率为100%。根尖孔正常的病例(A组)在根尖定位的误差明显小于有侧支根管的病例(B组)(p<0.001)。A组在±0.5毫米公差水平下的准确率为100%。