Fox R E, Redstone D
Am J Clin Pathol. 1976 Oct;66(4):658-66. doi: 10.1093/ajcp/66.4.658.
An investigation was carried out to examine why a glucose oxidase-peroxidase-orthodianisidine method for plasma glucose, without protein precipitation, gave low results for neonatal blood. The magnitude of the difference between the results with and without protein precipitation was examined in a clinical neonatal series, and in sera to which bilirubin, hemolysate, pure hemoglobin, and uric acid had been added. Systematic linear inhibition was demonstrated with bilirubin, and the results suggested that high concentrations of hemolysate and uric acid could also interfere. Use of alkaline protein precipitants eliminated the inhibition. Dextrostix test results for neonatal blood are compared with results of conventional glucose analyses and possible sources of discrepancy examined.
开展了一项调查,以研究为何一种用于血浆葡萄糖检测的葡萄糖氧化酶 - 过氧化物酶 - 邻联茴香胺法(不进行蛋白质沉淀)在检测新生儿血液时结果偏低。在一个临床新生儿系列以及添加了胆红素、溶血产物、纯血红蛋白和尿酸的血清中,检测了进行蛋白质沉淀和不进行蛋白质沉淀时结果的差异程度。结果表明胆红素存在系统性线性抑制作用,同时高浓度的溶血产物和尿酸也可能产生干扰。使用碱性蛋白质沉淀剂可消除这种抑制作用。将新生儿血液的葡萄糖试纸检测结果与传统葡萄糖分析结果进行了比较,并对可能存在差异的原因进行了研究。