Suppr超能文献

尿液中白蛋白和肌酐的仪器读取试纸条与视觉结果及定量方法的比较。

Comparison of instrument-read dipsticks for albumin and creatinine in urine with visual results and quantitative methods.

作者信息

Pugia M J, Lott J A, Luke K E, Shihabi Z K, Wians F H, Phillips L

机构信息

Diagnostics Business Group, Bayer Corporation, Elkhart, Indiana, USA.

出版信息

J Clin Lab Anal. 1998;12(5):280-4. doi: 10.1002/(sici)1098-2825(1998)12:5<280::aid-jcla5>3.0.co;2-a.

Abstract

Three hospital sites evaluated the Bayer two-pad urine dipstick as a screening test for microalbuminuria. One pad estimates albumin concentrations between 10 and 150 mg/L, and the second estimates creatinine values between 300 and 3,000 mg/L. The Boehringer Mannheim (BMD) Micral dipstick was also compared and evaluated. The accuracy of the dipsticks was judged by comparison with cuvet-based immunonephelometry for albumin and to standard rate-Jaffe methods for creatinine; these assays were well standardized and controlled and were assumed to give accurate values. Precision of these methods and that of the dipsticks was determined by multiple assays of control materials. Visual or instrument (Clinitek 50 or 100) evaluation of the Bayer or visual checks of the BMD albumin dipstick pad with patients' urines gave clinically acceptable accuracy. The albumin/creatinine ratio from the Bayer dipsticks gave better accuracy for albumin excretion than the albumin pads alone from either manufacturer. This ratio should permit making a good estimate of the 24-hr albumin excretion in a randomly collected urine.

摘要

三个医院站点对拜耳双垫尿试纸作为微量白蛋白尿的筛查测试进行了评估。一个试纸垫用于估计白蛋白浓度在10至150毫克/升之间,另一个试纸垫用于估计肌酐值在300至3000毫克/升之间。同时对罗氏(BMD)微量白蛋白尿试纸进行了比较和评估。通过与基于比色杯的白蛋白免疫比浊法以及肌酐的标准速率-贾菲法进行比较来判断试纸的准确性;这些检测方法经过了良好的标准化和控制,并假定能给出准确值。通过对对照材料进行多次检测来确定这些方法和试纸的精密度。对拜耳试纸进行视觉或仪器(Clinitek 50或100)评估,或对罗氏白蛋白试纸垫与患者尿液进行视觉检查,均获得了临床可接受的准确性。与单独使用任一制造商的白蛋白试纸垫相比,拜耳试纸的白蛋白/肌酐比值在白蛋白排泄方面具有更高的准确性。该比值应有助于对随机收集尿液中的24小时白蛋白排泄量做出良好估计。

相似文献

1
Comparison of instrument-read dipsticks for albumin and creatinine in urine with visual results and quantitative methods.
J Clin Lab Anal. 1998;12(5):280-4. doi: 10.1002/(sici)1098-2825(1998)12:5<280::aid-jcla5>3.0.co;2-a.
2
Comparison of urine dipsticks with quantitative methods for microalbuminuria.
Eur J Clin Chem Clin Biochem. 1997 Sep;35(9):693-700.
3
Multisite evaluation of a new dipstick for albumin, protein, and creatinine.
J Clin Lab Anal. 2001;15(5):231-5. doi: 10.1002/jcla.1032.
4
Accuracy of the spot urinary microalbumin:creatinine ratio and visual dipsticks in hypertensive pregnant women.
Eur J Obstet Gynecol Reprod Biol. 2009 Jun;144(2):146-8. doi: 10.1016/j.ejogrb.2009.03.010. Epub 2009 Apr 15.
7
Screening school children for albuminuria, proteinuria and occult blood with dipsticks.
Clin Chem Lab Med. 1999 Feb;37(2):149-57. doi: 10.1515/CCLM.1999.027.
9
Urinary protein and albumin excretion corrected by creatinine and specific gravity.
Clin Chim Acta. 2000 Apr;294(1-2):139-55. doi: 10.1016/s0009-8981(00)00181-9.
10
Diagnostic accuracy of urine dipsticks for detection of albuminuria in the general community.
Am J Kidney Dis. 2011 Jul;58(1):19-28. doi: 10.1053/j.ajkd.2010.12.026. Epub 2011 Mar 15.

引用本文的文献

1
Urinary strips for protein assays: easy to do but difficult to interpret!
J Nephrol. 2021 Apr;34(2):411-432. doi: 10.1007/s40620-020-00735-y. Epub 2020 Apr 23.
2
Using protein/creatinine ratios in random urine.
J Clin Lab Anal. 2005;19(4):160-6. doi: 10.1002/jcla.20071.
4
Measurement of the albumin content of urinary protein using dipsticks.
J Clin Lab Anal. 1999;13(5):246-50. doi: 10.1002/(SICI)1098-2825(1999)13:5&#x0003c;246::AID-JCLA10&#x0003e;3.0.CO;2-W.
5
High-sensitivity dye binding assay for albumin in urine.
J Clin Lab Anal. 1999;13(4):180-7. doi: 10.1002/(sici)1098-2825(1999)13:4<180::aid-jcla7>3.0.co;2-r.

本文引用的文献

1
Comparison of urine dipsticks with quantitative methods for microalbuminuria.
Eur J Clin Chem Clin Biochem. 1997 Sep;35(9):693-700.
2
Clinical significance of microalbuminuria and evaluation of the Micral-Test.
Clin Biochem. 1993 Aug;26(4):283-7. doi: 10.1016/0009-9120(93)90126-q.
3
Diurnal variations of blood pressure and microalbuminuria in essential hypertension.
Am J Hypertens. 1994 Jan;7(1):23-9. doi: 10.1093/ajh/7.1.23.
4
Factors related to the presence of microalbuminuria in essential hypertension.
Am J Hypertens. 1994 Sep;7(9 Pt 1):801-7. doi: 10.1093/ajh/7.9.801.
8
Prevention of diabetic renal disease with special reference to microalbuminuria.
Lancet. 1995 Oct 21;346(8982):1080-4. doi: 10.1016/s0140-6736(95)91747-0.
9
Microalbuminuria predicts clinical proteinuria and early mortality in maturity-onset diabetes.
N Engl J Med. 1984 Feb 9;310(6):356-60. doi: 10.1056/NEJM198402093100605.
10
Use of single voided urine samples to estimate quantitative proteinuria.
N Engl J Med. 1983 Dec 22;309(25):1543-6. doi: 10.1056/NEJM198312223092503.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验