Disconnect alarm failure in detection of common gas outlet disconnection.

作者信息

French R A, Kennedy R R

机构信息

Department of Anaesthesia, Christchurch Hospital, New Zealand.

出版信息

Anaesth Intensive Care. 1998 Dec;26(6):665-70. doi: 10.1177/0310057X9802600610.

Abstract

This laboratory study was prompted by two paediatric cases where low pressure alarms and capnography failed to detect common gas outlet disconnection when using a T-piece pump ventilator. A carbon dioxide producing model lung was ventilated using the Clare ventilator (a T-piece pump type ventilator) via an Ayres T-piece. The T-piece used has 3 mm diameter fresh gas tubing. Common gas outlet disconnects were carried out using varying fresh gas flows and ventilatory patterns. At fresh gas flows of less than 8 l/min, the low pressure alarm of the Clare ventilator failed to sound following common gas outlet disconnection. Oxygraphy showed a rapid rise in inspired oxygen. End-tidal CO2 rose, whilst inspired CO2 remained at zero. This experiment demonstrates that the Clare ventilator's low pressure alarm detects common gas outlet disconnection poorly when used with an Ayres T-piece with narrow fresh gas tubing. Graphical representation of airway oxygen content has merit as an alerting monitor for common gas outlet disconnection.

摘要

这项实验室研究是由两例儿科病例引发的,在使用T形管泵式呼吸机时,低压警报和二氧化碳监测未能检测到常见的气体出口断开情况。使用克莱尔呼吸机(一种T形管泵式呼吸机)通过艾尔斯T形管对一个产生二氧化碳的模拟肺进行通气。所使用的T形管有直径3毫米的新鲜气体管道。通过改变新鲜气体流量和通气模式进行常见气体出口断开操作。在新鲜气体流量低于8升/分钟时,克莱尔呼吸机的低压警报在常见气体出口断开后未响起。氧含量监测显示吸入氧迅速上升。呼气末二氧化碳上升,而吸入二氧化碳仍为零。该实验表明,当与具有狭窄新鲜气体管道的艾尔斯T形管一起使用时,克莱尔呼吸机的低压警报对常见气体出口断开的检测效果不佳。气道氧含量的图形表示作为常见气体出口断开的警报监测手段具有价值。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验