红外瞳孔测量法与对比法的一致性和可重复性。

Agreement and repeatability of infrared pupillometry and the comparison method.

作者信息

Wachler B S, Krueger R R

机构信息

Department of Ophthalmology, Jules Stein Eye Institute, UCLA School of Medicine, USA.

出版信息

Ophthalmology. 1999 Feb;106(2):319-23. doi: 10.1016/s0161-6420(99)90070-2.

Abstract

OBJECTIVE

To evaluate the accuracy and repeatability of the widely used comparison method of measuring pupil size.

DESIGN

Cross-sectional study.

PARTICIPANTS

Fourteen eyes of seven healthy myopic subjects were examined.

INTERVENTION

Two examiners made two repeated measures of pupil diameters of 14 eyes using Rosenbaum card comparison pupillometry and infrared pupillometry. Subjects fixated on a distant visual acuity chart, and pupils were measured under three luminance conditions. The agreement and inter-rater repeatability of both methods were determined.

MAIN OUTCOME MEASURES

Outcomes were pupil diameters, limits of agreement, and coefficient of repeatability of two examiners.

RESULTS

The mean difference between the two techniques ranged from 0.3 to 0.5 mm. The limits of agreement within two standard deviations ranged from 2.4 to 2.8 mm. Coefficient of repeatability ranged from 0.6 to 1.4 mm for infrared pupillometry and 1.0 to 1.2 mm for Rosenbaum pupillometry. Inter-rater repeatability of Rosenbaum pupillometry was consistently pupil diameter biased. Pupil diameters measured with the Rosenbaum method were consistently larger than diameters measured with the infrared technique for both examiners under all luminance conditions.

CONCLUSIONS

Results indicate that although the mean difference in techniques was small, the range of the agreement between the Rosenbaum and the infrared techniques was large. The Rosenbaum method consistently overestimated pupil diameters and was subject to inter-rater repeatability bias. Rosenbaum pupillometry may not be appropriate when accurate pupil measurements are required. The results have implications for many clinical trials in ophthalmology, including those evaluating refractive surgery that use Rosenbaum comparison pupillometry.

摘要

目的

评估广泛使用的测量瞳孔大小比较方法的准确性和可重复性。

设计

横断面研究。

参与者

对7名健康近视受试者的14只眼睛进行了检查。

干预措施

两名检查者使用罗森鲍姆卡片比较式瞳孔测量法和红外瞳孔测量法对14只眼睛的瞳孔直径进行了两次重复测量。受试者注视远处视力表,在三种亮度条件下测量瞳孔。确定了两种方法的一致性和检查者间的可重复性。

主要观察指标

观察指标为瞳孔直径、一致性界限和两名检查者的可重复性系数。

结果

两种技术之间的平均差异在0.3至0.5毫米之间。两个标准差内的一致性界限在2.4至2.8毫米之间。红外瞳孔测量法的可重复性系数在0.6至1.4毫米之间,罗森鲍姆瞳孔测量法的可重复性系数在1.0至1.2毫米之间。罗森鲍姆瞳孔测量法的检查者间可重复性始终存在瞳孔直径偏差。在所有亮度条件下,两名检查者使用罗森鲍姆方法测量的瞳孔直径始终大于使用红外技术测量的直径。

结论

结果表明,尽管技术之间的平均差异很小,但罗森鲍姆法与红外技术之间的一致性范围很大。罗森鲍姆方法始终高估瞳孔直径,并存在检查者间可重复性偏差。当需要精确测量瞳孔时,罗森鲍姆瞳孔测量法可能不合适。这些结果对许多眼科临床试验有影响,包括那些评估使用罗森鲍姆比较式瞳孔测量法的屈光手术的试验。

文献检索