Krieglstein G K, Waller W K
Albrecht Von Graefes Arch Klin Exp Ophthalmol. 1975;194(1):11-6. doi: 10.1007/BF00408271.
For evaluation of hand-applanation tonometry comparative measurements of intraocular pressure have been carried out using the Goldmann applanation tonometer, the Draeger and Perkins hand-applanation tonometers and the Schiötz indentation tonometer. A series of 52 eyes has been tested. The statistical analysis was based on the Goldmann tonometry as a reference system. The hand-applanation tonometers were found to be in satisfactory agreement with the Goldmann tonometer, whereas the Schiötz tonometer gave significantly low values in the pressure range considered.
为评估手持压平眼压计,使用戈德曼压平眼压计、德尔格和珀金斯手持压平眼压计以及施茨压陷眼压计进行了眼压的对比测量。对一系列52只眼睛进行了测试。统计分析以戈德曼眼压测量法作为参照系统。结果发现,手持压平眼压计与戈德曼眼压计的测量结果吻合度良好,而在上述眼压范围内,施茨眼压计给出的值明显偏低。