Suppr超能文献

使用Humphrey 2000光学相干断层扫描技术测量黄斑厚度的学期间重复性。

Intersession repeatability of macular thickness measurements with the Humphrey 2000 OCT.

作者信息

Koozekanani D, Roberts C, Katz S E, Herderick E E

机构信息

Biomedical Engineering Center, College of Medicine and Public Health, The Ohio State University, Columbus, USA.

出版信息

Invest Ophthalmol Vis Sci. 2000 May;41(6):1486-91.

Abstract

PURPOSE

This study was designed to determine intersession repeatability of measurements of macular thickness made with a commercially available optical coherence tomography (OCT) system. The images that can be routinely acquired with the commercial instrument differ significantly in quality from the images in the literature, which have mostly been acquired on prototype systems.

METHODS

Multiple OCT images of the nasal macula were obtained from the right eye during three independent measuring sessions, using the Humphrey 2000 OCT system (Humphrey, San Leandro, CA). Twenty-six volunteers with no history of ocular disease participated in this investigation. Eyes in all subjects were undilated during scan acquisition. Scans were horizontal, 3 mm long, and through the fovea. Five scans were used from each session, for a total of 15 scans per subject. Retinal boundaries were automatically detected using custom software. Statistical software was used to calculate intersession and intrasession repeatability. Manual correction was performed on the automatically detected boundaries, and a second analysis was performed using these boundaries.

RESULTS

When no manual correction of boundaries was performed, there were no significant effects between different sessions (P = 0.529) or between different scans within the same session (P = 0.509). Average retinal thickness was found to be 274 +/- 17 microm for a 1-mm long region 0.75 mm from the fovea. Individual scan averages differed from overall patient averages by 0 +/- 4.3 microm (99% confidence interval, 11.2 microm).

CONCLUSIONS

OCT measurements of macular thickness made with the Humphrey 2000 OCT system are repeatable over different sessions with an expected variation of less than 11 microm (99% confidence interval).

摘要

目的

本研究旨在确定使用商用光学相干断层扫描(OCT)系统测量黄斑厚度的不同测量时段之间的可重复性。商用仪器常规获取的图像在质量上与文献中的图像有显著差异,文献中的图像大多是在原型系统上获取的。

方法

使用Humphrey 2000 OCT系统(Humphrey,加利福尼亚州圣莱安德罗),在三个独立的测量时段从右眼获取鼻侧黄斑的多个OCT图像。26名无眼部疾病史的志愿者参与了本研究。扫描采集过程中所有受试者的眼睛均未散瞳。扫描为水平方向,长3 mm,穿过中央凹。每个时段使用5次扫描,每位受试者共15次扫描。使用定制软件自动检测视网膜边界。使用统计软件计算不同测量时段之间和同一测量时段内的可重复性。对自动检测的边界进行手动校正,并使用这些边界进行第二次分析。

结果

当未对边界进行手动校正时,不同测量时段之间(P = 0.529)或同一测量时段内不同扫描之间(P = 0.509)均无显著影响。发现距中央凹0.75 mm处1 mm长区域的平均视网膜厚度为274±17微米。个体扫描平均值与患者总体平均值的差异为0±4.3微米(99%置信区间,11.2微米)。

结论

使用Humphrey 2000 OCT系统测量黄斑厚度在不同测量时段具有可重复性,预期变化小于11微米(99%置信区间)。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验