Swanson W H, Felius J, Birch D G
Retina Foundation of the Southwest, Dallas, Texas, USA.
Ophthalmology. 2000 Oct;107(10):1950-4. doi: 10.1016/s0161-6420(00)00356-0.
To determine the effect of stimulus size on sensitivity of patients with retinitis pigmentosa (RP) as measured by automated static perimetry.
Comparative case series.
Thirty-nine patients with RP and a control group of 10 healthy volunteers.
Automated static perimetry (full threshold programs 24-2 or 30-2) was performed twice on one eye of each participant using stimulus sizes III (0.43 degrees diameter) and V (1.72 degrees diameter). Data from the same 50 test locations were used from each field.
At each location, for each participant, the size effect was computed as the difference (in decibels) in sensitivities for sizes V and III, and the average sensitivity was computed as the mean of sensitivities for the two sizes.
For both patient and control groups, the size effect was negatively correlated with average sensitivity (r(2) > 0.124; P: < 0.001). The mean size effect was significantly greater for the patient group than for the control group: 8.6 (+/- 3.6) dB versus 5. 4 (+/- 2.2) dB (t = 18.0; P: < 0.001). The percentage of abnormal locations (more than 8 dB below mean normal) tended to be lower for size V than for size III, with a mean of 67% for size V versus 95% for size III. The percentage of absolute defects was also lower for size V than for size III, with a mean of 35% for size V versus 54% for size III.
In damaged regions of the visual fields of patients with RP, increase in stimulus size from III to V can produce abnormally large increases in perimetric sensitivity. Size III may be more useful than size V for detection of field abnormality, whereas size V may be more useful than size III for observing progression of advanced RP.
通过自动静态视野计测定刺激大小对视网膜色素变性(RP)患者敏感度的影响。
比较病例系列研究。
39例RP患者及10名健康志愿者作为对照组。
对每位参与者的一只眼睛使用刺激大小III(直径0.43度)和V(直径1.72度)进行两次自动静态视野计检查(全阈值程序24 - 2或30 - 2)。每个视野均采用相同的50个测试位点的数据。
在每个位点,对每位参与者计算大小效应,即大小V和III敏感度差值(分贝),并计算平均敏感度,即两种大小敏感度的均值。
对于患者组和对照组,大小效应均与平均敏感度呈负相关(r² > 0.124;P < 0.001)。患者组的平均大小效应显著大于对照组:8.6(±3.6)分贝对5.4(±2.2)分贝(t = 18.0;P < 0.001)。大小V异常位点(比正常均值低8分贝以上)的百分比往往低于大小III,大小V平均为67%,大小III为95%。大小V绝对缺损的百分比也低于大小III,大小V平均为35%,大小III为54%。
在RP患者视野的受损区域,刺激大小从III增加到V可使视野敏感度出现异常大幅增加。大小III在检测视野异常方面可能比大小V更有用,而大小V在观察晚期RP进展方面可能比大小III更有用。