通过在色素性视网膜炎中剔除低敏感性的结果,改善静态自动视野计的重测变异性。

Improvements in Test-Retest Variability of Static Automated Perimetry by Censoring Results With Low Sensitivity in Retinitis Pigmentosa.

机构信息

Department of Ophthalmology, Stein Eye Institute, University of California, Los Angeles, Los Angeles, CA, USA.

College of Optometry, Nova Southeastern University, Fort Lauderdale, FL, USA.

出版信息

Transl Vis Sci Technol. 2020 Nov 20;9(12):26. doi: 10.1167/tvst.9.12.26. eCollection 2020 Nov.

Abstract

PURPOSE

We evaluated whether omitting (censoring) points in more severely damaged visual field areas can reduce test-retest variability of static automated perimetry (SAP) in retinitis pigmentosa (RP), as variability creates a significant challenge when monitoring for changes.

METHODS

Cohort 1 included 27 eyes in 16 RP subjects with visual acuity (VA) ranging from 20/20 to 20/70 who completed Humphrey 10-2 size III SAP, once per visit at three visits. Cohort 2 included 15 eyes in nine RP subjects with VA ≤ 20/60 who completed Humphrey 30-2 size V SAP, twice per visit at three visits. Variability was assessed using 95% coefficient of repeatability (CR) calculations for uncensored (all threshold values and data points included) and censored data.

RESULTS

In cohort 1, the uncensored between-visit 95% CR was 11.6 decibels (dB); censoring locations with threshold values of <8 to 20 dB resulted in 31% to 53% reductions in the 95% CR. For cohort 2, uncensored 95% CRs were 8.7 and 8.0 dB for within- and between-visit variability, respectively; censoring <8 to 17 dB resulted in 15% to 41% and 15% to 43% reductions in within-visit and between-visit 95% CRs, respectively. For both cohorts, censoring at higher values yielded slightly less variability, at the expense of discarding data from a greater number of eyes and test locations.

CONCLUSIONS

For 20/20 to 20/70 VA tested with size III stimuli, censoring lower sensitivity values results in substantially lower test-retest variability, which may help detect true changes for locations without severe baseline loss.

TRANSLATIONAL RELEVANCE

A rule of thumb for clinical practices using SAP to monitor RP is that longitudinal losses of >9 dB for individual test locations with initial values ≥ 9 dB are likely to be real and meaningful, as they exceed typical variability.

摘要

目的

我们评估了在视网膜色素变性(RP)的静态自动视野检查(SAP)中,是否可以通过省略(删失)更严重受损的视野区域的点来降低测试-重测变异性,因为变异性在监测变化时会带来重大挑战。

方法

队列 1 纳入了 16 名 RP 患者的 27 只眼,这些患者的视力(VA)范围从 20/20 到 20/70,他们在三次就诊中每次就诊时完成 Humphrey 10-2 大小 III SAP。队列 2 纳入了 9 名 RP 患者的 15 只眼,这些患者的 VA≤20/60,他们在三次就诊中每次就诊时完成 Humphrey 30-2 大小 V SAP。使用 95%重复性系数(CR)计算评估未删失(包括所有阈值值和数据点)和删失数据的变异性。

结果

在队列 1 中,未删失的两次就诊之间的 95% CR 为 11.6 分贝(dB);对阈值值<8 至 20 dB 的位置进行删失,可使 95% CR 降低 31%至 53%。对于队列 2,未删失的内和两次就诊之间的 95% CR 分别为 8.7 和 8.0 dB;对<8 至 17 dB 的位置进行删失,可使内和两次就诊之间的 95% CR 分别降低 15%至 41%和 15%至 43%。对于两个队列,较高值的删失会导致变异性略低,但代价是从更多的眼睛和测试位置丢弃数据。

结论

对于 VA 为 20/20 至 20/70 的患者进行 III 大小刺激测试,删失较低的灵敏度值可显著降低测试-重测变异性,这可能有助于检测无严重基线损失的位置的真实变化。

临床意义

使用 SAP 监测 RP 的临床实践的一条经验法则是,对于初始值≥9 dB 的个别测试位置,纵向损失>9 dB 可能是真实且有意义的,因为它们超过了典型的变异性。