Briggs C, Harrison P, Grant D, Staves J, MacHin S J
Department of Haematology, University College London Hospital, London, UK.
Clin Lab Haematol. 2000 Dec;22(6):345-50. doi: 10.1046/j.1365-2257.2000.00330.x.
The XE 2100 (Sysmex Corporation) is a cell counter that furthers the technology of fluorescent flow cytometry developed from the earlier range of Sysmex analysers. The new diagnostic features are a nucleated red cell count (NRBC), the ability to measure platelets by impedance as well as an 'optical' platelet count using a fluorescence dye and an immature granulocyte (IG) count. The NRBC count was highly correlated (r = 0.97) with the manual reference count. For counts below 100 x 109/l the 'optical' method and the immunocount gave good a correlation (r = 0.97) optical and impedance counts were also well correlated (r = 0.89). The use of the 'optical' platelet count significantly improves the reliability of low platelet counts. The IG count correlated with visual counts (r = 0.81) and allows the detection of immature cells at an earlier stage in the laboratory process. The introduction of fluorescent flow cytometric analysis allows extended quantification of additional cell populations and so potentially improves screening and monitoring of various pathological conditions.
XE 2100(希森美康公司)是一款血细胞分析仪,它在希森美康早期分析仪系列所采用的荧光流式细胞术技术基础上进行了改进。新的诊断功能包括有核红细胞计数(NRBC)、通过阻抗法测量血小板的能力以及使用荧光染料的“光学”血小板计数和未成熟粒细胞(IG)计数。NRBC计数与手工参考计数高度相关(r = 0.97)。对于低于100×10⁹/L的计数,“光学”方法与免疫计数具有良好的相关性(r = 0.97),光学计数和阻抗计数也具有良好的相关性(r = 0.89)。“光学”血小板计数的使用显著提高了低血小板计数的可靠性。IG计数与目视计数相关(r = 0.81),并能够在实验室检测过程的早期阶段检测到未成熟细胞。荧光流式细胞术分析的引入使得对其他细胞群体进行扩展定量成为可能,从而潜在地改善了对各种病理状况的筛查和监测。