使用1mol/L钆布醇的X射线数字减影血管造影术:与碘化造影剂对比的猪研究结果

X-ray digital subtraction angiography with 1 mol/L gadobutrol: results from a comparative porcine study With iodinated contrast agents.

作者信息

Kalinowski Marc, Kress Oliver, Wels Thomas, Alfke Heiko, Klose Klaus Jochen, Wagner Hans-Joachim

机构信息

Department of Diagnostic Radiology, Philipps-University Hospital, Marburg, Germany.

出版信息

Invest Radiol. 2002 May;37(5):254-62. doi: 10.1097/00004424-200205000-00003.

Abstract

RATIONALE AND OBJECTIVES

To evaluate prospectively diagnostic accuracy of 1 mol/L gadobutrol as a contrast agent for intraarterial x-ray digital subtraction angiography (DSA) in comparison to iodinated, nonionic contrast media and 0.5 mol/L gadolinium-DTPA.

METHODS

Flush arteriograms (ascending, descending, abdominal aorta, iliac, and femoral arteries) and selective angiograms (carotid, renal, and visceral arteries) were obtained from bilateral femoral arterial access (5 F sheaths) in 10 domestic pigs (70 kg body weight). Digital subtracted angiograms were obtained during injection of undiluted 1 mol/L gadobutrol, 300 mg I/mL iopromide, or 0.5 mol/L gadopentetate. Injection parameters (volume and velocity) were similar for all three contrast agents. In paired arteries, two different contrast media were used during the same angiographic run. Diagnostic quality and accuracy of the angiograms were evaluated on a three-step scale by three independent blinded investigators.

RESULTS

Sufficient nonselective angiographic images were obtained in 90% of cases using iodinated contrast material. Gadobutrol achieved sufficient nonselective angiograms in 64%. Selective angiograms were sufficient in 98% using iodinated contrast material, 90% using 1 mol/L Gadobutrol and 48% using 0.5 mol/L Gd-DTPA. Adverse reactions to any of the used contrast agents were not noted.

CONCLUSION

One mol/L Gadobutrol solution allows x-ray digital subtraction angiography with a diagnostic accuracy equivalent to 300 mg/mL iodinated contrast media, if selective injections are performed. Flush aortograms are of inferior image quality to iodinated contrast material.

摘要

原理与目的

前瞻性评估1 mol/L钆布醇作为动脉内X线数字减影血管造影(DSA)造影剂相对于碘化非离子型造影剂和0.5 mol/L钆喷酸葡胺的诊断准确性。

方法

从10头家猪(体重70 kg)的双侧股动脉入路(5F鞘)获取冲洗动脉造影(升主动脉、降主动脉、腹主动脉、髂动脉和股动脉)和选择性血管造影(颈动脉、肾动脉和内脏动脉)。在注射未稀释的1 mol/L钆布醇、300 mg I/mL碘普罗胺或0.5 mol/L钆喷酸葡胺期间获取数字减影血管造影。三种造影剂的注射参数(体积和速度)相似。在配对动脉中,在同一次血管造影过程中使用两种不同的造影剂。由三名独立的盲法研究者按照三步量表评估血管造影的诊断质量和准确性。

结果

使用碘化造影剂时,90%的病例获得了足够的非选择性血管造影图像。钆布醇在64%的病例中获得了足够的非选择性血管造影图像。使用碘化造影剂时,98%的选择性血管造影图像足够,使用1 mol/L钆布醇时为90%,使用0.5 mol/L钆喷酸葡胺时为48%。未观察到对任何所用造影剂的不良反应。

结论

如果进行选择性注射,1 mol/L钆布醇溶液可进行X线数字减影血管造影,诊断准确性与300 mg/mL碘化造影剂相当。冲洗主动脉造影的图像质量低于碘化造影剂。