Nelesen Richard A, Dimsdale Joel E
Department of Psychiatry, University of California, San Diego, La Jolla, 92093, USA.
Blood Press Monit. 2002 Oct;7(5):259-63. doi: 10.1097/00126097-200210000-00002.
Non-invasive continuous blood pressure monitoring has, for the past decade, been performed using the Finapres monitor. Radial tonometry is a new methodology for measuring non-invasive beat-to-beat blood pressure. This technique has been evaluated for other applications, but its use in cardiovascular reactivity testing has not yet been reported.
A convenience sample was used to compare a radial tonometric device (Colin Pilot) with intermittent brachial blood pressure measurement (Critikon Dinamap) during reactivity testing. A second comparison was made between an Ohmeda Finapres and a Dinamap.
Twenty subject-testing sessions during a reactivity protocol were evaluated for blood pressure determination. Separate groups of subjects were used for Pilot versus Dinamap comparisons and Finapres versus Dinamap comparisons. Blood pressure signals from 15 subjects with Pilot and 15 subjects with Finapres data were evaluated for artifacts. Signals were also compared in a subject with Raynaud's phenomenon.
Both devices tracked blood pressure well. The Finapres significantly overestimated systolic blood pressure ( P < 0.001). The correlations between the Pilot and Dinamap were consistently higher than those between the Finapres and Dinamap. There were more artifacts (3.2%) with the Finapres than with the Pilot (0.2%). In the subject with Raynaud's phenomenon, the Finapres overestimated blood pressure by more than 40 mmHg and showed more than 75% of artifacts. On the other hand, the Pilot tracked blood pressure in the same subject without error.
Radial tonometric continuous beat-to-beat blood pressure monitoring with the Colin Pilot was associated with a low artifact rating and high accuracy of blood pressure measurement during reactivity testing.
在过去十年中,一直使用Finapres监护仪进行无创连续血压监测。桡动脉压力测量法是一种测量无创逐搏血压的新方法。该技术已在其他应用中得到评估,但尚未见其在心血管反应性测试中的应用报道。
在反应性测试期间,采用便利样本将桡动脉压力测量装置(Colin Pilot)与间歇性肱动脉血压测量(Critikon Dinamap)进行比较。还对Ohmeda Finapres和Dinamap进行了第二次比较。
在反应性方案的20次受试者测试过程中,对血压测定进行评估。将不同组的受试者用于Pilot与Dinamap的比较以及Finapres与Dinamap的比较。对15名有Pilot数据的受试者和15名有Finapres数据的受试者的血压信号进行伪差评估。还对一名患有雷诺现象的受试者的信号进行了比较。
两种装置对血压的跟踪效果都很好。Finapres显著高估收缩压(P<0.001)。Pilot与Dinamap之间的相关性始终高于Finapres与Dinamap之间的相关性。Finapres的伪差(3.2%)比Pilot(0.2%)更多。在患有雷诺现象的受试者中,Finapres高估血压超过40 mmHg,且伪差超过75%。另一方面,Pilot在同一受试者中对血压的跟踪没有误差。
在反应性测试期间,使用Colin Pilot进行桡动脉压力测量连续逐搏血压监测具有低伪差率和高血压测量准确性。