在接受脊髓麻醉的产科患者中对一种连续无创血压监测仪的评估。

Evaluation of a continuous noninvasive blood pressure monitor in obstetric patients undergoing spinal anesthesia.

作者信息

Epstein R H, Kaplan S, Leighton B L, Norris M C, DeSimone C A

机构信息

Department of Anesthesiology, Jefferson Medical College, Thomas Jefferson University, Philadelphia, PA 19107.

出版信息

J Clin Monit. 1989 Jul;5(3):157-63. doi: 10.1007/BF01627447.

Abstract

A noninvasive blood pressure monitor (Finapres) that continuously displays the arterial waveform using the Penaz methodology has recently been introduced into clinical practice. We compared this device with an automated oscillometric blood pressure monitor (Dinamap 1846SX) in 20 patients during spinal anesthesia for nonemergency cesarean section according to a procedure suggested by the Association for the Advancement of Medical Instrumentation. After administration of the spinal anesthetic, the Finapres monitor produced systolic, mean, and diastolic pressure measurements greater than those of the Dinamap monitor (6.6 +/- 12.5, 3.3 +/- 10.4, and 7.2 +/- 9.8 mm Hg, respectively). In most patients, the Finapres measurements were similar to those determined by the Dinamap; however, in 4 patients, mean systolic differences were greater than 20 mm Hg. These patients did not differ from the others in age, height, weight, or baseline blood pressure, and the pressure values recorded by the Finapres monitor were substantially higher than those measured by auscultation in the labor room. In 30% of the patients, the offset between Dinamap and Finapres blood pressure measurements changed markedly over the course of the surgical procedure. The Finapres monitor occasionally stopped working and had to be restarted. In 1 patient (not included in this analysis), the Dinamap monitor was unable to determine the blood pressure due to patient shivering; this did not appear to interfere with the Finapres. We conclude that the Finapres monitor does not consistently provide blood pressure information equivalent to that of the Dinamap in obstetric patients undergoing spinal anesthesia.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

一种采用Penaz方法持续显示动脉波形的无创血压监测仪(Finapres)最近已引入临床实践。我们按照医疗仪器促进协会建议的程序,在20例非急诊剖宫产脊髓麻醉期间,将该设备与自动示波血压监测仪(Dinamap 1846SX)进行了比较。脊髓麻醉给药后,Finapres监测仪测得的收缩压、平均压和舒张压高于Dinamap监测仪(分别为6.6±12.5、3.3±10.4和7.2±9.8毫米汞柱)。在大多数患者中,Finapres测量值与Dinamap测定值相似;然而,在4例患者中,平均收缩压差异大于20毫米汞柱。这些患者在年龄、身高、体重或基线血压方面与其他患者无差异,Finapres监测仪记录的血压值明显高于产房听诊测量值。在30%的患者中,Dinamap和Finapres血压测量值之间的偏差在手术过程中显著变化。Finapres监测仪偶尔会停止工作,必须重新启动。在1例患者(未纳入本分析)中,由于患者颤抖,Dinamap监测仪无法测定血压;这似乎并未干扰Finapres。我们得出结论,在接受脊髓麻醉的产科患者中,Finapres监测仪不能始终提供与Dinamap相当的血压信息。(摘要截短于250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验