Cuff bladder width and blood pressure measurement in children and adolescents.

作者信息

Gómez-Marín O, Prineas R J, Råstam L

机构信息

Department of Epidemiology and Public Health, University of Miami School of Medicine, FL 33101.

出版信息

J Hypertens. 1992 Oct;10(10):1235-41.

PMID:1335006
Abstract

OBJECTIVE

To estimate the independent contribution of the blood pressure cuff bladder width: mid-upper arm circumference (CW:AC) ratio to the variability in systolic blood pressure (SBP), diastolic blood pressure phase 4 (DBP4) and diastolic blood pressure phase 5 (DBP5) in children and adolescents, and its impact upon the estimate of prevalence of high blood pressure derived at screening.

DESIGN

Use of cuffs with a CW:AC ratio below 40% causes overestimation of the true blood pressure level. With higher CW:AC ratios the level is underestimated. National recommendations for adults state that the CW:AC ratio is optimal at approximately 40%, but no consistent recommendations exist for children.

METHODS

We measured SBP and diastolic blood pressure phases 4 and 5 in 811 boys and 771 girls aged 10-17 years (93% participation rate). In each subject the blood pressure was measured with a cuff conforming to the 40% rule (recommended cuff), with one smaller (mean decrease in the CW:AC ratio:7.4%), and with one larger (mean increase in the CW:AC ratio:10.2%). Cuffs were used in random order.

RESULTS

Smaller cuffs gave significantly higher blood pressure readings. With the larger cuff there were significant mean decreases. The differences were independent of the blood pressure level obtained with the recommended cuff. The impact upon the estimate of the prevalence of high blood pressure was substantial.

CONCLUSIONS

Selection of the proper cuff size is important for children and adolescents, in order to avoid both over- and underdiagnosis of hypertension.

摘要

目的

评估血压袖带气囊宽度与上臂中段周长之比(CW:AC)对儿童和青少年收缩压(SBP)、舒张压第4期(DBP4)和舒张压第5期(DBP5)变异性的独立影响,及其对筛查时高血压患病率估计值的影响。

设计

使用CW:AC比低于40%的袖带会高估真实血压水平。CW:AC比越高,血压水平被低估。针对成年人的国家建议指出,CW:AC比约为40%时最为理想,但对于儿童尚无一致的建议。

方法

我们测量了811名10至17岁男孩和771名女孩(参与率93%)的SBP、DBP4和DBP5。在每个受试者身上,分别使用符合40%规则的袖带(推荐袖带)、一个较小的袖带(CW:AC比平均降低7.4%)和一个较大的袖带(CW:AC比平均增加10.2%)测量血压。袖带按随机顺序使用。

结果

较小的袖带测得的血压读数显著更高。使用较大的袖带时,平均血压有显著下降。这些差异与使用推荐袖带测得的血压水平无关。对高血压患病率估计值的影响很大。

结论

为避免高血压的过度诊断和漏诊,为儿童和青少年选择合适的袖带尺寸很重要。