Ansari-Lari M Ali, Kickler Thomas S, Borowitz Michael J
Department of Pathology, Johns Hopkins Medical Institutions, 401 N Broadway, Room Weinberg 2335, Baltimore, MD 21231-2410, USA.
Am J Clin Pathol. 2003 Nov;120(5):795-9. doi: 10.1309/LT30-BV9U-JJV9-CFHQ.
We determined the usefulness of immature granulocyte measurement as a predictor of infection or positive blood culture and compared the results with total WBC count and absolute neutrophil count (ANC). Blood samples from 102 infected and 69 noninfected patients were analyzed using the Sysmex XE-2100 automated blood cell counter (Sysmex, Kobe, Japan). The percentage of immature granulocytes was significantly higher in infected than in noninfected patients and in patients with positive than patients with negative blood cultures. Receiver operating characteristic curves showed that the percentage of immature granulocytes was a better predictor of infection than the WBC count and comparable to the ANC. Automated immature granulocyte measurements reflect a biologically and clinically relevant phenomenon but are not sensitive enough to be used as screening assays for prediction of infection or bacteremia. However, although infrequently encountered, a percentage of immature granulocytes of more than 3 was a very specific predictor of sepsis and might help expedite microbiologic laboratory evaluation of a subset of patients.
我们确定了未成熟粒细胞测量作为感染或血培养阳性预测指标的有效性,并将结果与白细胞总数和绝对中性粒细胞计数(ANC)进行比较。使用Sysmex XE - 2100全自动血细胞分析仪(Sysmex,日本神户)对102例感染患者和69例未感染患者的血样进行分析。感染患者中未成熟粒细胞的百分比显著高于未感染患者,血培养阳性患者中未成熟粒细胞的百分比显著高于血培养阴性患者。受试者工作特征曲线显示,未成熟粒细胞百分比比白细胞计数是更好的感染预测指标,与ANC相当。自动未成熟粒细胞测量反映了一种生物学和临床相关现象,但灵敏度不足以用作预测感染或菌血症的筛查检测。然而,虽然不常见,但未成熟粒细胞百分比超过3是脓毒症的一个非常特异的预测指标,可能有助于加快对一部分患者的微生物实验室评估。