Panjabi M, Chang D, Dvorák J
Department of Orthopaedics and Rehabilitation, Yale University School of Medicine, New Haven, Connecticut.
Spine (Phila Pa 1976). 1992 Feb;17(2):200-5. doi: 10.1097/00007632-199202000-00014.
A pair of functional radiographs, taken at each end of the range of motion, are used to determine spinal motions. Graphic construction and computer-assisted methods are available for the radiographic analysis. The later provides many more motion parameters. A study of lumbar spine lateral radiographs was conducted to determine errors in the motion parameters due to spinal level, radiographic quality, and errors in the two digitizing instruments. Significant differences were found in the errors due to the two digitizers when the same radiographic pair was redigitized several times. There were only minimal differences, however, between the digitizers when the radiographic films were remarked and redigitized. The error ranges (2 x SD) for the motion parameters were 1) rotation = +/- 1.25 degrees; 2) translation of the inferior posterior vertebral body corner = +/- 0.86 degrees; and 3) coordinates for the center of rotation = +/- 4.3 mm. Both the spinal level and radiographic quality affected the magnitude of errors in all motion parameters.
在运动范围的两端各拍摄一对功能X线片,用于确定脊柱运动。有图形构建和计算机辅助方法可用于X线分析。后者可提供更多的运动参数。进行了一项腰椎侧位X线片研究,以确定由于脊柱节段、X线质量以及两种数字化仪器的误差导致的运动参数误差。当同一对X线片被多次重新数字化时,发现两种数字化仪导致的误差存在显著差异。然而,当对X线片重新标注并重新数字化时,两种数字化仪之间只有极小的差异。运动参数的误差范围(2倍标准差)为:1)旋转=±1.25度;2)下后椎体角的平移=±0.86度;3)旋转中心的坐标=±4.3毫米。脊柱节段和X线质量均影响所有运动参数的误差大小。