It has been shown in several in-vitro studies that IVUS images correlate well with histology and angioscopy. The existing data on intravascular imaging of large-diameter elastic arteries are scant. The aim of this study was to compare the IVUS images of aorta with the reference method--histological section by evaluating the accuracy of measurements performed at corresponding aortic segments.
15 human aortic specimens were derived from post-mortem examination and in-vitro IVUS imaging was performed, including measurements of vessel diameters and cross-sectional areas in three specified areas. Corresponding regions of interest obtained with histology were subsequently evaluated and parameters matched to those by IVUS were calculated.
Total of 45 pairs of measurement were compared. Mean vessel internal diameter on IVUS was 14.4 mm (SD 2.7 mm) and intima thickness was 0.6 mm (SD 0.8 mm). The corresponding aortic internal diameter observed with histology was 13.2 (SD 2.6 mm) and intima thickness--0.3 mm (SD 0.03). An overall agreement between IVUS and histology specimens was high, yelling correlation ratio of 0.99. The highest level of agreement was observed for external vessel area validation--correlation ratio 0.98. The largest discrepancy was observed for intima and media thickness, with correlation ratios of 0.61 and 0.72, respectively.
Our data suggest that examination of aorta with intravascular ultrasound is feasible and may provide additional information in assessment of cardiovascular pathology. More complete risk stratification, which includes IVUS examination of aorta, is an important factor to determine the necessity for early preventive treatment, especially in patients with normal coronary arteries.
多项体外研究表明,血管内超声(IVUS)图像与组织学及血管镜检查结果相关性良好。关于大口径弹性动脉血管内成像的现有数据较少。本研究旨在通过评估在相应主动脉节段进行测量的准确性,将主动脉的IVUS图像与参考方法——组织学切片进行比较。
从尸检中获取15例人类主动脉标本,并进行体外IVUS成像,包括在三个特定区域测量血管直径和横截面积。随后评估通过组织学获得的相应感兴趣区域,并计算与IVUS测量参数相匹配的参数。
共比较了45对测量值。IVUS测得的平均血管内径为14.4毫米(标准差2.7毫米),内膜厚度为0.6毫米(标准差0.8毫米)。组织学观察到的相应主动脉内径为13.2(标准差2.6毫米),内膜厚度为0.3毫米(标准差0.03)。IVUS与组织学标本之间的总体一致性较高,相关系数为0.99。在血管外膜面积验证方面观察到最高水平的一致性——相关系数为0.98。在内膜和中膜厚度方面观察到最大差异,相关系数分别为0.61和0.72。
我们的数据表明,使用血管内超声检查主动脉是可行的,并且在评估心血管病理学方面可能提供额外信息。更全面的风险分层,包括对主动脉进行IVUS检查,是确定早期预防性治疗必要性的重要因素,尤其是对于冠状动脉正常的患者。