Tonnu P-A, Ho T, Sharma K, White E, Bunce C, Garway-Heath D
Glaucoma Research Unit, Moorfields Eye Hospital, London EC1V 2PD, UK.
Br J Ophthalmol. 2005 Jul;89(7):847-50. doi: 10.1136/bjo.2004.056614.
To compare the inter-method agreement in intraocular pressure (IOP) measurements made with four different tonometric methods.
IOP was measured with the Goldmann applanation tonometer (GAT), Tono-Pen XL, ocular blood flow tonograph (OBF), and Canon TX-10 non-contact tonometer (NCT) in a randomised order in one eye of each of 105 patients with ocular hypertension or glaucoma. Three measurements were made with each method, and by each of two independent GAT observers. GAT interobserver and tonometer inter-method agreement was assessed by the Bland-Altman method. The outcome measures were 95% limits of agreement for IOP measurements between GAT observers and between tonometric methods, and 95% confidence intervals for intra-session repeated measurements.
The mean differences (bias) in IOP measurements were 0.4 mm Hg between GAT observers, and 0.6 mm Hg, 0.1 mm Hg, and 0.7 mm Hg between GAT and Tono-Pen, OBF, and NCT, respectively. The 95% limits of agreement were smallest (bias +/-2.6 mm Hg) between GAT observers, and larger for agreement between the GAT and the Tono-Pen, OBF, and NCT (bias +/-6.7, +/-5.5, and +/-4.8 mm Hg, respectively). The OBF and NCT significantly underestimated GAT measurements at lower IOP and overestimated these at higher IOP. The repeatability coefficients for intra-session repeated measurement for each method were +/-2.2 mm Hg and +/-2.5 mm Hg for the GAT, +/-4.3 mm Hg for the Tono-Pen, +/-3.7 mm Hg for the OBF, and +/-3.2 mm Hg for the NCT.
There was good interobserver agreement with the GAT and moderate agreement between the NCT and GAT. The differences between the GAT and OBF and between the GAT and Tono-Pen probably preclude the OBF and Tono-Pen from routine clinical use as objective methods to measure IOP in normal adult eyes.
比较四种不同眼压测量方法测量眼内压(IOP)的方法间一致性。
对105例高眼压症或青光眼患者的单眼,随机采用Goldmann压平眼压计(GAT)、Tono-Pen XL眼压计、眼血流眼压描记仪(OBF)和佳能TX-10非接触眼压计(NCT)测量眼压。每种方法测量3次,且由两名独立的GAT观察者分别进行测量。采用Bland-Altman方法评估GAT观察者间和眼压计方法间的一致性。观察指标为GAT观察者间和眼压计方法间IOP测量的95%一致性界限,以及同一测量时段重复测量的95%置信区间。
GAT观察者间IOP测量的平均差异(偏差)为0.4 mmHg,GAT与Tono-Pen、OBF和NCT之间的平均差异分别为0.6 mmHg、0.1 mmHg和0.7 mmHg。GAT观察者间的95%一致性界限最小(偏差±2.6 mmHg),GAT与Tono-Pen、OBF和NCT之间的一致性界限较大(分别为偏差±6.7、±5.5和±4.8 mmHg)。在较低眼压时,OBF和NCT显著低估GAT测量值,在较高眼压时则高估这些测量值。每种方法同一测量时段重复测量的重复性系数,GAT为±2.2 mmHg和±2.5 mmHg,Tono-Pen为±4.3 mmHg,OBF为±3.7 mmHg,NCT为±3.2 mmHg。
GAT观察者间一致性良好,NCT与GAT之间一致性中等。GAT与OBF以及GAT与Tono-Pen之间的差异可能使OBF和Tono-Pen无法作为测量正常成人眼压的客观方法用于常规临床。