Keeler Pulsair 3000与Goldmann压平眼压计的临床比较。

Clinical comparison of the Keeler Pulsair 3000 with Goldmann applanation tonometry.

作者信息

Parker V A, Herrtage J, Sarkies N J

机构信息

Department of Ophthalmology, Addenbrooke's Hospital Trust, Cambridge, UK.

出版信息

Br J Ophthalmol. 2001 Nov;85(11):1303-4. doi: 10.1136/bjo.85.11.1303.

Abstract

AIM

To confirm the accuracy of the Pulsair 3000 before introducing the instrument into clinical practice.

METHOD

A masked study by two experienced tonometrists comparing the mean of the Goldmann intraocular pressure (IOP) readings against the Pulsair 3000 reading (average of four puffs). Results of 150 eyes were compared with an IOP range of 10 mm Hg-44 mm Hg.

RESULTS

Correlation between the two Goldmann tonometry results was 0.9830 with a standard deviation of 1.1085 mm Hg. Correlation between the mean of the two Goldmann readings against the Pulsair 3000 reading was 0.982 with a standard deviation of 1.1179 mm Hg. Bland-Altman analysis confirms a satisfactory outcome.

CONCLUSION

The Pulsair 3000 provides an accurate and objective method of measuring IOP with many advantages over traditional Goldmann tonometry.

摘要

目的

在将Pulsair 3000仪器引入临床实践之前,确认其准确性。

方法

由两名经验丰富的眼压测量师进行一项盲法研究,比较Goldmann眼压(IOP)读数的平均值与Pulsair 3000读数(四次吹气的平均值)。将150只眼睛的结果与10毫米汞柱至44毫米汞柱的眼压范围进行比较。

结果

两次Goldmann眼压测量结果之间的相关性为0.9830,标准差为1.1085毫米汞柱。两次Goldmann读数的平均值与Pulsair 3000读数之间的相关性为0.982,标准差为1.1179毫米汞柱。Bland-Altman分析证实结果令人满意。

结论

Pulsair 3000提供了一种准确、客观的眼压测量方法,与传统的Goldmann眼压测量法相比有许多优点。

相似文献

8
Evaluation of the Tono-Pen and the Pulsair tonometers.眼压笔和脉冲式眼压计的评估。
Am J Ophthalmol. 1990 Jun 15;109(6):716-20. doi: 10.1016/s0002-9394(14)72442-0.

引用本文的文献

7
Tonometers-which one should I use?眼压计——我应该用哪一种?
Eye (Lond). 2018 May;32(5):931-937. doi: 10.1038/s41433-018-0040-4. Epub 2018 Feb 19.
8
Detecting IOP Fluctuations in Glaucoma Patients.检测青光眼患者的眼压波动
Open Ophthalmol J. 2016 Feb 29;10:44-55. doi: 10.2174/1874364101610010044. eCollection 2016.

本文引用的文献

4
Effects of repeated tonometry: genuine and sham measurements.反复眼压测量的效果:真实测量与假测量
Acta Ophthalmol (Copenh). 1972;50(4):574-82. doi: 10.1111/j.1755-3768.1972.tb05987.x.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验