Huang Ronnie Y C, Lam Andrew K C, Chan Rufina, Young Shui-Ming
Department of Optometry and Radiography, The Hong Kong Polytechnic University, Kowloon, Hong Kong, China.
Ophthalmic Physiol Opt. 2005 Sep;25(5):441-5. doi: 10.1111/j.1475-1313.2005.00321.x.
To compare the Orbscan central corneal thickness (CCT) values before and after intraocular pressure (IOP) measurement with Goldmann applanation tonometry in young normal adults.
Fifty-three eyes of 53 subjects who were free from any ocular problems were studied. The measurements included optical pachometry by the Orbscan II system, followed by Goldmann applanation tonometry, and finally Orbscan optical pachometry again. Standard Goldmann technique was adopted with the application of one drop of 0.4% benoxinate and fluorescein sodium prior to the measurement. The same operator was involved in the Orbscan pachometry before and after Goldmann tonometry. Another investigator was responsible for Goldmann tonometry. Three readings were obtained in each case, and the means were used for analysis.
The mean IOP of the sample was 14.6 +/- 2.6 mmHg. There was no significant difference (paired t-test: p = 0.50) in the mean CCT values before and after the Goldmann tonometry (551 +/- 32 and 552 +/- 35 microm respectively).
Standard Goldmann applanation tonometry does not affect the Orbscan CCT values. Measurement of CCT for the adjustment of true IOP can be carried out either before or after Goldmann tonometry.
比较年轻正常成年人使用Goldmann压平眼压计测量眼压前后Orbscan中央角膜厚度(CCT)值。
对53名无任何眼部问题的受试者的53只眼进行研究。测量包括使用Orbscan II系统进行光学测厚,随后进行Goldmann压平眼压测量,最后再次进行Orbscan光学测厚。测量前应用一滴0.4%丁氧卡因和荧光素钠,采用标准Goldmann技术。Goldmann眼压测量前后进行Orbscan测厚由同一操作者完成。另一位研究者负责Goldmann眼压测量。每种情况均获取三次读数,并采用平均值进行分析。
样本的平均眼压为14.6±2.6 mmHg。Goldmann眼压测量前后的平均CCT值无显著差异(配对t检验:p = 0.50)(分别为551±32和552±35微米)。
标准Goldmann压平眼压测量不影响Orbscan CCT值。用于调整真实眼压的CCT测量可在Goldmann眼压测量之前或之后进行。