Effect of central corneal thickness on Goldmann applanation tonometry measures - a different result with different pachymeters.

作者信息

Doughty Michael J, Jonuscheit Sven

机构信息

Department of Vision Sciences, Glasgow-Caledonian University, Glasgow G4 OBA, UK.

出版信息

Graefes Arch Clin Exp Ophthalmol. 2007 Nov;245(11):1603-10. doi: 10.1007/s00417-007-0601-x. Epub 2007 May 24.

Abstract

PURPOSE

To assess whether the use of Orbscan pachymetry, as an alternative to ultrasound pachymetry, had an effect on the IOP-CCT relationship for Goldmann tonometry in normal human eyes.

PATIENTS AND METHODS

The right eye of 50 healthy subjects, aged 19 to 82 years, were assessed by Orbscan (three scans), and then central corneal thickness (CCT) measured by ultrasound pachymetry (three readings) and Goldmann tonometry performed (three readings).

RESULTS

Goldmann tonometry indicated average IOP values from 9.7 to 20 mm Hg (group mean 14.3 +/- 2.5 mm Hg) that were highly highly correlated with ultrasound pachymetry (r = 0.608, p < 0.001). Orbscan central zone pachymetry (2-mm sample zone) yielded higher values than ultrasound, averaging 0.591 +/- 0.044 mm compared to 0.523 +/- 0.037 mm (p < 0.001), with the differences between the instruments being proportional to the average thickness (p < 0.001; r = 0.461). Notwithstanding, the Goldmann tonometry values were still highly correlated with the Orbscan central zone thickness data (r = 0.595, p < 0.001); but, due to the difference in the thickness data generated by the two pachymeters, the absolute slope of the IOP-CCT relationship was slightly less if Orbscan measures were used, with or without use of the acoustic factor.

CONCLUSION

Orbscan measures of the thickness of a central corneal zone can be used to assess the impact of central corneal thickness on Goldmann tonometry data.