非洲裔美国人中动态轮廓眼压计与Goldmann压平眼压计的比较。

Comparison of dynamic contour tonometry and goldmann applanation tonometry in African American subjects.

作者信息

Medeiros Felipe A, Sample Pamela A, Weinreb Robert N

机构信息

Hamilton Glaucoma Center and Department of Ophthalmology, University of California, San Diego, La Jolla, California 92093-0946, USA.

出版信息

Ophthalmology. 2007 Apr;114(4):658-65. doi: 10.1016/j.ophtha.2006.06.065. Epub 2006 Nov 30.

Abstract

PURPOSE

To evaluate the relationship between intraocular pressure (IOP) measurements obtained by dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT) in African Americans and to assess whether these measures were influenced by ocular parameters including corneal thickness, corneal curvature, and axial length.

DESIGN

Observational clinical study.

PARTICIPANTS

The study included 176 eyes of 94 African Americans.

METHODS

All participants underwent IOP evaluation with DCT and GAT, as well as measurements of central corneal thickness (CCT), corneal curvature, corneal astigmatism, spherical equivalent, and axial length.

MAIN OUTCOME MEASURES

Univariate and multiple regression analyses were used to evaluate the associations between IOP (as measured with DCT and GAT) and CCT, corneal curvature, corneal astigmatism, spherical equivalent, axial length, and age. Bland-Altman plots were used to evaluate the agreement between IOP measurements obtained by DCT and GAT.

RESULTS

In multiple regression analysis, GAT IOP measurements were significantly associated with CCT (P<0.001) and age (P = 0.003), whereas DCT IOP measurements were associated only with age (P = 0.027). The difference between DCT and GAT IOP measurements was significantly influenced by corneal thickness (P<0.001) and ocular pulse amplitude (OPA; P = 0.004). Patients with thick corneas tended to have higher GAT IOP measurements compared with DCT, whereas in patients with thin corneas, GAT IOP measurements tended to be lower than DCT measurements. Larger values of OPA were associated with a decrease in the difference between DCT and GAT measurements, whereas smaller values of OPA were associated with an increase in the difference between DCT and GAT measurements.

CONCLUSIONS

Dynamic contour tonometry measurements in African Americans seem to provide an estimate of IOP that is less influenced by corneal properties than those provided by GAT.

摘要

目的

评估动态轮廓眼压计(DCT)和Goldmann压平眼压计(GAT)所测眼压(IOP)在非裔美国人中的关系,并评估这些测量值是否受包括角膜厚度、角膜曲率和眼轴长度在内的眼部参数影响。

设计

观察性临床研究。

参与者

该研究纳入了94名非裔美国人的176只眼。

方法

所有参与者均接受了DCT和GAT眼压评估,以及中央角膜厚度(CCT)、角膜曲率、角膜散光、等效球镜度和眼轴长度测量。

主要观察指标

采用单因素和多因素回归分析评估眼压(通过DCT和GAT测量)与CCT、角膜曲率、角膜散光、等效球镜度、眼轴长度和年龄之间的关联。采用Bland-Altman图评估DCT和GAT所测眼压的一致性。

结果

在多因素回归分析中,GAT所测眼压与CCT(P<0.001)和年龄(P = 0.003)显著相关,而DCT所测眼压仅与年龄相关(P = 0.027)。DCT和GAT所测眼压的差异受角膜厚度(P<0.001)和眼动脉搏动幅度(OPA;P = 0.004)显著影响。角膜厚的患者与DCT相比,GAT所测眼压往往更高,而角膜薄的患者中,GAT所测眼压往往低于DCT测量值。OPA值越大,DCT和GAT测量值之间的差异越小,而OPA值越小,DCT和GAT测量值之间的差异越大。

结论

非裔美国人中,动态轮廓眼压计测量似乎提供了一种比GAT所测眼压受角膜特性影响更小的眼压估计值。