Non-invasive measurement of cardiac output by thoracic electrical bioimpedance: a study of reproducibility and comparison with thermodilution.

作者信息

Jewkes C, Sear J W, Verhoeff F, Sanders D J, Foëx P

机构信息

Nuffield Department of Anaesthetics, University of Oxford, Radcliffe Infirmary.

出版信息

Br J Anaesth. 1991 Dec;67(6):788-94. doi: 10.1093/bja/67.6.788.

Abstract

The performance and reproducibility of the BoMED NCCOM3 thoracic electrical bioimpedance cardiograph (TEB) has been evaluated in volunteers and patients. In resting supine volunteers, we determined the coefficient of variability over short time periods (30 min) and over several days, and examined the effects of differences in electrode type and electrode placement. The mean (range) intra-subject coefficients of variation (CV) for thoracic fluid index (TFI) and stroke volume (SV) were 1.0% (0.4-1.8%) and 4.7% (2.1-8.5%), respectively over a 30-min period. The corresponding CV were 5.6% (2.3-10.9%) and 10.9% (6.1-14.8%) for measurements made at rest on four separate occasions. Use of different electrode types (RedDot and Medicotest) resulted in differences in TFI (P less than 0.01), but not in mean values for SV or cardiac output (Q); their use in individual subjects revealed differences of up to 20% in SV and Q. Alterations in electrode placement by 5 cm in the horizontal and diagonal planes produced no significant changes in TFI, SV or Q; changes in the longitudinal plane produced a graded change. Increases of 5 cm and 10 cm in thoracic length produced mean increases in TFI of 9.8% and 39.8%, respectively, and mean decreases in Q of 8.4% and 16.7% and SV of 7.5% and 15.8%. TEB measurements of Q and SV were compared with thermodilution (TD) in 16 intensive care patients. Mean (SEM) Q by TEB was 5.63 (1.10) litre min-1 compared with TD 4.38 (0.72) litre min-1 (P less than 0.01).(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验