Dziezyc J, Millichamp N J, Smith W B
Department of Small Animal Medicine and Surgery, College of Veterinary Medicine, Texas A&M University, College Station 77843.
J Am Vet Med Assoc. 1992 Aug 1;201(3):430-3.
Two Mackay-Marg tonometers and 2 Tono-Pen tonometers were evaluated in eyes in which intraocular pressure (IOP) had been altered and measured by use of a manometer. Eyes of anesthetized dogs and enucleated horse eyes were used. Compared with the manometer, none of the tonometers accurately measured IOP over the range between 0 and 100 mm of Hg. However at manometer measurements from 0 to 30 mm of Hg, several of the tonometers accurately measured IOP. In addition, significant differences were observed when the measurement accuracy of one tonometer was compared with that of another, especially at high IOP. Coefficient of determination (r2) values for a linear model ranged from 0.979 to 0.991 in dogs, and from 0.982 to 0.996 in horse eyes.