The influence of different speech processor and hearing aid settings on speech perception outcomes in electric acoustic stimulation patients.

作者信息

Vermeire Katrien, Anderson Ilona, Flynn Mark, Van de Heyning Paul

机构信息

Dept ENT Universitar Ziekenhuis, Antwerp, Belgium.

出版信息

Ear Hear. 2008 Jan;29(1):76-86. doi: 10.1097/AUD.0b013e31815d6326.

Abstract

OBJECTIVE

Electric acoustic stimulation (EAS) is an increasingly popular means of treating individuals with a steeply sloping mid-to-high frequency hearing loss, who traditionally do not benefit from hearing instruments. These persons often have too much residual hearing to be considered for a cochlear implant. Several studies have demonstrated the ability both to preserve the remaining low-frequency hearing in these individuals, and to provide significant benefit through combining a cochlear implant with a hearing aid to amplify the same ear. These improvements in performance have been especially noted in noise. Often overlooked is that these outcomes may be influenced by the fitting parameters of both the cochlear implant and the hearing aid.

DESIGN

This study assessed four EAS subjects, with a minimum of 1 month's EAS use, on eight different fitting parameters. Sentence testing in different noise levels (+15, +10, and +5 dB SPL) was conducted. Subjects also evaluated each condition using a visual analogue scale.

RESULTS

Results demonstrated that a reduced overlap of cochlear implant and hearing aid amplification produced best results across listening conditions.

CONCLUSIONS

The hearing aid should be fit to a patient-specific modified audiogram at least up to the point where low-frequency hearing is not measurable. The cochlear implant should be fit from a higher frequency point than is standard in patients without residual hearing in the implanted ear, to provide reduced overlap with the amplification provided by the hearing aid. Therefore, a small amount of overlap between the frequency ranges used by the hearing aid and the cochlear implant seems beneficial.

摘要

目的

电声刺激(EAS)是一种越来越流行的治疗方法,用于治疗中高频听力损失呈陡坡状的个体,这些个体传统上无法从助听器中获益。这些人通常有过多的残余听力,不适合植入人工耳蜗。多项研究表明,该方法既能保留这些个体剩余的低频听力,又能通过将人工耳蜗与助听器结合来放大同一耳朵,从而带来显著益处。这些性能改善在噪声环境中尤为明显。常常被忽视的是,这些结果可能会受到人工耳蜗和助听器两者的调试参数的影响。

设计

本研究评估了四名使用电声刺激至少1个月的受试者,对八个不同的调试参数进行测试。在不同噪声水平(+15、+10和+5 dB SPL)下进行句子测试。受试者还使用视觉模拟量表对每种情况进行评估。

结果

结果表明,人工耳蜗和助听器放大范围的重叠减少在各种聆听条件下产生了最佳效果。

结论

助听器应根据患者特定的修正听力图进行调试,至少要达到低频听力无法测量的程度。人工耳蜗的调试频率点应高于植入耳无残余听力的患者的标准频率点,以减少与助听器放大范围的重叠。因此,助听器和人工耳蜗使用的频率范围之间有少量重叠似乎是有益的。