Suppr超能文献

3T下的局部激发黑血成像:在颈动脉壁的应用

Local excitation black blood imaging at 3T: application to the carotid artery wall.

作者信息

Bornstedt Axel, Bernhardt Peter, Hombach Vinzenz, Kamenz Joachim, Spiess Jochen, Subgang Anne, Rasche Volker

机构信息

Experimental Cardiovascular Imaging, Department of Internal Medicine II, University Hospital Ulm, University of Ulm, Robert-Koch-Strasse 8, Ulm, Germany.

出版信息

Magn Reson Med. 2008 May;59(5):1207-11. doi: 10.1002/mrm.21590.

Abstract

A novel approach for imaging large sections of the carotid artery wall at isotropic spatial resolution is presented. Local excitation by means of 2D excitation pulses was combined with a diffusion-prepared segmented steady-state black-blood gradient echo technique enabling the assessment of the carotid arterial wall over a range of up to 15 cm. The carotid arteries of five healthy volunteers were imaged with the proposed technique. Signal-to-noise ratio (SNR), wall-lumen contrast-to-noise ratio (CNR), and vessel dimensions were assessed and compared to conventional excitation techniques. In all experiments black-blood contrast could be realized over the covered carotid arteries with similar SNR and CNR as the conventional technique covering the region of the bulbus only. The proposed technique enables the time-efficient coverage of the carotid arteries without compromising wall-lumen CNR and geometrical accuracy. Furthermore, the proposed technique appears to be less sensitive to motion and swallowing artifacts due to the local character of the excitation.

摘要

本文提出了一种以各向同性空间分辨率对颈动脉壁大切片进行成像的新方法。通过二维激发脉冲进行局部激发,并结合扩散准备的分段稳态黑血梯度回波技术,能够在长达15厘米的范围内评估颈动脉壁。使用所提出的技术对五名健康志愿者的颈动脉进行成像。评估了信噪比(SNR)、壁腔对比噪声比(CNR)和血管尺寸,并与传统激发技术进行了比较。在所有实验中,在所覆盖的颈动脉上均可实现黑血对比,其SNR和CNR与仅覆盖球部区域的传统技术相似。所提出的技术能够在不影响壁腔CNR和几何精度的情况下高效地覆盖颈动脉。此外,由于激发的局部特性,所提出的技术似乎对运动和吞咽伪影不太敏感。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验