Mårtensson Mattias, Winter Reidar, Cederlund Kerstin, Ripsweden Jonaz, Mir-Akbari Habib, Nowak Jacek, Brodin Lars-Ake
School for Technology and Health, Royal Institute of Technology, Flemingsberg, Stockholm, Sweden.
Cardiovasc Ultrasound. 2008 Jun 4;6:26. doi: 10.1186/1476-7120-6-26.
To compare the accuracy of simplified 3-dimensional (3-D) echocardiography vs. multi-slice computed tomography (MSCT) software for the quantification of left ventricular (LV) volumes.
Three-D echocardiography (3-planes approach) and MSCT-CardIQ software were calibrated by measuring known volumes of 10 phantoms designed to closely mimic blood-endocardium interface. Subsequently, LV volumes were measured with both the methods in 9 patients referred routinely for coronary angiography and the agreement between the measurements was evaluated.
Simplified 3D-echocardiography provided higher degree of agreement between the measured and true phantom volumes (mean difference 0 +/- 1 ml, variation range +4 to -4 ml) than MSCT software (mean difference 6 +/- 5 ml; variation range +22 to -10 ml). The agreement between LV measurements in the patients was considerably poorer, with significantly larger volumes produced by MSCT (mean difference -23 +/- 40 ml, variation between +93 and -138 ml).
Simplified 3-D echocardiography provides more accurate assessment of phantom volumes than MSCT-CardIQ software. The discrepancy between the results of LV measurements with the two methods is even greater and does not warrant their interchangeable diagnostic use.