傅里叶域与时域光学相干断层扫描技术在视神经带状萎缩检测中的比较

Comparison of Fourier-domain and time-domain optical coherence tomography in the detection of band atrophy of the optic nerve.

作者信息

Costa-Cunha Luciana V F, Cunha Leonardo P, Malta Roberto F S, Monteiro Mário L R

机构信息

Neuro-ophthalmology Service, Hospital das Clínicas of the University of São Paulo Medical School, São Paulo, Brazil.

出版信息

Am J Ophthalmol. 2009 Jan;147(1):56-63.e2. doi: 10.1016/j.ajo.2008.07.020. Epub 2008 Sep 6.

Abstract

PURPOSE

To compare the ability of Fourier-domain (FD) optical coherence tomography (3D OCT-1000; Topcon, Tokyo, Japan) and time-domain (TD) OCT (Stratus; Carl Zeiss Meditec Inc, Dublin, California, USA) to detect axonal loss in eyes with band atrophy (BA) of the optic nerve.

DESIGN

Cross-sectional study.

METHODS

Thirty-six eyes from 36 patients with BA and temporal visual field (VF) defect from chiasmal compression and 36 normal eyes were studied. Subjects were submitted to standard automated perimetry and macular and retinal nerve fiber layer (RNFL) measurements were taken using 3D OCT-1000 and Stratus OCT. Receiver operating characteristic (ROC) curves were calculated for each parameter. Spearman correlation coefficients were obtained to evaluate the relationship between RNFL and macular thickness parameters and severity of VF loss. Measurements from the two devices were compared.

RESULTS

Regardless of OCT device, all RNFL and macular thickness parameters were significantly lower in eyes with BA compared with normal eyes, but no statistically significant difference was found with regard to the area under the ROC curve. Structure-function relationships were also similar for the two devices. In both groups, RNFL and macular thickness measurements were generally and in some cases significantly smaller with 3D OCT-1000 than with Stratus OCT.

CONCLUSIONS

The introduction of FD technology did not lead to better discrimination ability for detecting BA of the optic nerve compared with TD technology when using the software currently provided by the manufacturer. 3D OCT-1000 FD OCT RNFL and macular measurements were generally smaller than TD Stratus OCT measurements. Investigators should be aware of this fact when comparing measurements obtained with these two devices.

摘要

目的

比较傅里叶域(FD)光学相干断层扫描(3D OCT - 1000;日本东京拓普康公司)和时域(TD)光学相干断层扫描(Stratus;美国加利福尼亚州都柏林卡尔蔡司医疗技术公司)检测视神经带状萎缩(BA)患者眼部轴突损失的能力。

设计

横断面研究。

方法

对36例因视交叉受压导致BA和颞侧视野(VF)缺损患者的36只眼以及36只正常眼进行研究。受试者接受标准自动视野检查,并使用3D OCT - 1000和Stratus OCT测量黄斑和视网膜神经纤维层(RNFL)。计算每个参数的受试者操作特征(ROC)曲线。获得Spearman相关系数以评估RNFL和黄斑厚度参数与VF损失严重程度之间的关系。比较两种设备的测量结果。

结果

无论使用哪种OCT设备,与正常眼相比,BA患者眼中所有RNFL和黄斑厚度参数均显著降低,但ROC曲线下面积无统计学显著差异。两种设备的结构 - 功能关系也相似。在两组中,3D OCT - 1000测量的RNFL和黄斑厚度通常且在某些情况下显著小于Stratus OCT。

结论

与TD技术相比,在使用制造商目前提供的软件时,FD技术的引入并未在检测视神经BA方面带来更好的辨别能力。3D OCT - 1000 FD OCT的RNFL和黄斑测量值通常小于TD Stratus OCT测量值。研究人员在比较这两种设备获得的测量结果时应注意这一事实。