A clinical evaluation of rapid automatic noninvasive blood pressure determination with the Ohmeda 2120 "return-to-flow" method.

作者信息

Block F E, Fletcher M V, Morris T J, Dzwonczyk R

机构信息

Department of Anesthesiology, Indian River Memorial Hospital, Vero Beach, FL.

出版信息

J Clin Monit. 1991 Jul;7(3):241-4. doi: 10.1007/BF01619267.

Abstract

Most commercial noninvasive blood pressure devices use the oscillometric method for determination of blood pressure. The Ohmeda 2120 noninvasive blood pressure monitor uses the oscillometric technique, but it also includes a "Sys Stat" mode for rapid determination of systolic blood pressure up to ten times per minute. This Sys Stat mode uses the "return-to-flow" method, in which the pressure on an upper arm cuff decreases in small steps until a finger sensor detects the first pulse of blood. Data from 16 patients monitored with the Ohmeda 2120 unit and with an arterial cannula were analyzed. The line of regression was Sys Stat = 0.85 (arterial) + 9.49. The Ohmeda 2120 monitor tended to underestimate the arterial pressure, particularly at higher pressures. Several factors could have contributed to this underestimation. Even so, the accuracy is sufficient for clinical use under most circumstances. More importantly, in an urgent situation, the Sys Stat mode can provide a reading of systolic blood pressure within seconds.

摘要

大多数商用无创血压测量设备采用示波法来测定血压。欧姆龙2120无创血压监测仪采用示波技术,但它还包括一种“Sys Stat”模式,可每分钟快速测定收缩压达十次。这种Sys Stat模式采用“血流恢复”法,即上臂袖带的压力以小步幅下降,直到手指传感器检测到第一个血流脉冲。对16名使用欧姆龙2120设备并进行动脉插管监测的患者的数据进行了分析。回归方程为Sys Stat = 0.85(动脉压)+ 9.49。欧姆龙2120监测仪往往会低估动脉压,尤其是在较高压力时。有几个因素可能导致了这种低估。即便如此,在大多数情况下,其准确性足以满足临床使用。更重要的是,在紧急情况下,Sys Stat模式可在数秒内提供收缩压读数。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验