Krengli Marco, Gaiano Simone, Mones Eleonora, Ballarè Andrea, Beldì Debora, Bolchini Cesare, Loi Gianfranco
Department of Radiotherapy, University Hospital Maggiore della Carità, Novara, Italy.
Radiat Oncol. 2009 Feb 22;4:9. doi: 10.1186/1748-717X-4-9.
The reproducibility of patient setup for radiotherapy is based on various methods including external markers, X-rays with planar or computerized image acquisition, and, more recently, surface matching imaging. We analyzed the setup reproducibility of 16 patients affected by prostate cancer who underwent conformal radiotherapy with curative intent by using a surface image registration system.
We analyzed the setup reproducibility of 16 patients affected by prostate cancer candidates for conformal radiotherapy by using a surface image registration system. At the initial setup, EPID images were compared with DRRs and a reference 3D surface image was obtained by the AlignRT system (Vision RT, London, UK). Surface images were acquired prior to every subsequent setup procedure. EPID acquisition was repeated when errors > 5 mm were reported.
The mean random and systematic errors were 1.2 +/- 2.3 mm and 0.3 +/- 3.0 mm along the X axis, 0.0 +/- 1.4 mm and 0.5 +/- 2.0 mm along the Y axis, and 2.0 +/- 1.8 mm and -0.7 +/- 2.4 mm along the Z axis respectively. The positioning error detected by AlignRT along the 3 axes X, Y, and Z exceeded the value of 5 mm in 14.1%, 2.0%, and 5.1% measurements and the value of 3 mm in 36.9%, 13.6% and 27.8% measurements, respectively. Correlation factors calculated by linear regression between the errors measured by AlignRT and EPID ranged from 0.77 to 0.92 with a mean of 0.85 and SD of 0.13. The setup measurements by surface imaging are highly reproducible and correlate with the setup errors detected by EPID.
Surface image registration system appears to be a simple, fast, non-invasive, and reproducible method to analyze the set-up alignment in 3DCRT of prostate cancer patients.
放射治疗中患者摆位的可重复性基于多种方法,包括外部标记、具有平面或计算机图像采集功能的X射线,以及最近的表面匹配成像。我们使用表面图像配准系统分析了16例接受根治性适形放疗的前列腺癌患者的摆位可重复性。
我们使用表面图像配准系统分析了16例适合适形放疗的前列腺癌患者的摆位可重复性。在初始摆位时,将电子射野影像装置(EPID)图像与数字重建放射影像(DRR)进行比较,并通过AlignRT系统(英国伦敦Vision RT公司)获得参考三维表面图像。在每次后续摆位程序之前采集表面图像。当报告误差>5 mm时重复进行EPID采集。
沿X轴的平均随机误差和系统误差分别为1.2±2.3 mm和0.3±3.0 mm,沿Y轴分别为0.0±1.4 mm和0.5±2.0 mm,沿Z轴分别为2.0±1.8 mm和-0.7±2.4 mm。AlignRT检测到的沿X、Y和Z三个轴的定位误差在测量中超过5 mm的值分别为14.1%、2.0%和5.1%,超过3 mm的值分别为36.9%、13.6%和27.8%。通过线性回归计算的AlignRT和EPID测量误差之间的相关系数范围为0.77至0.92,平均值为0.85,标准差为0.13。表面成像的摆位测量具有高度可重复性,并且与EPID检测到的摆位误差相关。
表面图像配准系统似乎是一种简单、快速、非侵入性且可重复的方法,用于分析前列腺癌患者三维适形放疗中的摆位对准情况。