Radiation Oncology Unit, Department of Clinical and Biological Sciences, University of Turin, S. Luigi Gonzaga Hospital, Regione Gonzole 10, 10043, Orbassano, Torino, Italy.
Radiol Med. 2012 Dec;117(8):1419-28. doi: 10.1007/s11547-012-0823-9. Epub 2012 May 14.
Accurate patient setup is a prerequisite for conformal radiotherapy (3D-CRT) and is based on various methods, including surface imaging systems. To evaluate the validity of a surface imaging system (AlignRT), we analysed setup reproducibility of a cohort of patients.
Nineteen patients affected by prostate adenocarcinoma were enrolled in this study. We acquired 653 surface images and 99 digital portal images (DPI). Setup errors were found by matching surface images with computed tomography (CT) and DPI images.
The setup errors from the threshold of 5 mm detected by AlignRT along the Y, Z and X axes occurred in 47.4%, 42.1% and 5.3% of patients, respectively. For the threshold of 3 mm, shifts along the Y, Z and X axes were observed in 68%, 69% and 10%, respectively. Comparing AlignRT and DPI, we found a statistically significant difference in the detection of shifts along the Y and Z axes. For a threshold ≥ 5 mm, the two systems provided corresponding setup errors along the Y and Z axes, whereas along the X axis, the threshold was not necessary.
AlignRT is an accurate technique for setup in 3D-CRT prostate cancer patients, especially along the lateral direction.
精确的患者摆位是适形放疗(3D-CRT)的前提,其基于多种方法,包括体表成像系统。为了评估体表成像系统(AlignRT)的准确性,我们分析了一组患者的摆位重复性。
本研究纳入了 19 例前列腺腺癌患者。我们共采集了 653 次体表图像和 99 次数字射野影像(DPI)。通过将体表图像与 CT 图像和 DPI 图像进行匹配,来发现摆位误差。
AlignRT 检测到的 5mm 阈值下沿 Y、Z 和 X 轴的摆位误差分别发生在 47.4%、42.1%和 5.3%的患者中。对于 3mm 的阈值,沿 Y、Z 和 X 轴的移动分别发生在 68%、69%和 10%的患者中。与 DPI 相比,AlignRT 在 Y 和 Z 轴上的移位检测中存在统计学差异。对于≥5mm 的阈值,两个系统沿 Y 和 Z 轴提供了相应的摆位误差,而在 X 轴上,该阈值则无必要。
AlignRT 是 3D-CRT 前列腺癌患者摆位的一种精确技术,尤其是在侧方方向。