Deantonio Letizia, Masini Laura, Loi Gianfranco, Gambaro Giuseppina, Bolchini Cesare, Krengli Marco
Department of Radiotherapy, University Hospital "Maggiore della Carità", Novara, Italy.
Department of Medical Physics, University Hospital "Maggiore della Carità", Novara, Italy.
Rep Pract Oncol Radiother. 2011 Apr 8;16(3):77-81. doi: 10.1016/j.rpor.2011.02.003. eCollection 2011.
To investigate the clinical application of a technique for patient set-up verification in breast cancer radiotherapy based on a 3D surface image registration system.
Accurate and reproducible patient set-up is a prerequisite to correctly deliver fractionated radiotherapy. Various approaches are available to verify and correct patient setup for 3D image acquisition in a radiation treatment room.
The study analyzed the setup reproducibility of 15 patients affected by breast cancer and candidates for conformal radiotherapy by using the AlignRT system (VisionRT, London, UK). At the initial setup, electronic portal imaging device (EPID) images were compared with Digitally Reconstructed Radiographs (DRRs) and a reference three-dimensional (3D) surface image was obtained by AlignRT. Surface images were acquired prior to every subsequent setup procedure. The systematic and random errors along longitudinal and vertical directions were measured and compared for the two systems.
The procedure for surface registration, image acquisition and comparison with the reference image took less than 1 min on average. The T test for systematic error showed no significant difference between the 2 verification systems along the longitudinal (p = 0.69) and vertical (p = 0.67) axes. The T-test for random error showed a significant difference between the 2 systems along the vertical axis (p = 0.05).
AlignRT is fast, simple, non-invasive and seems to be reliable in detecting patient setup errors. Our results suggest that it could be used to assess the setup reproducibility for breast cancer patients.
研究基于三维表面图像配准系统的乳腺癌放疗患者摆位验证技术的临床应用。
准确且可重复的患者摆位是正确实施分次放疗的前提。在放射治疗室中,有多种方法可用于验证和纠正患者在三维图像采集时的摆位。
本研究使用AlignRT系统(英国伦敦VisionRT公司)分析了15例乳腺癌患者及适形放疗候选患者的摆位重复性。在初始摆位时,将电子射野影像装置(EPID)图像与数字重建放射影像(DRR)进行比较,并通过AlignRT获得参考三维(3D)表面图像。在每次后续摆位程序之前采集表面图像。测量并比较了两个系统在纵向和垂直方向上的系统误差和随机误差。
表面配准、图像采集以及与参考图像比较的过程平均耗时不到1分钟。系统误差的T检验显示,两个验证系统在纵向(p = 0.69)和垂直(p = 0.67)轴上无显著差异。随机误差的T检验显示,两个系统在垂直轴上存在显著差异(p = 0.05)。
AlignRT快速、简单、无创,在检测患者摆位误差方面似乎可靠。我们的结果表明,它可用于评估乳腺癌患者的摆位重复性。