Abstract

The QuickVue Influenza A+B Test (Quidel) was used to test nasal swab specimens obtained from persons with influenza-like illness in 3 different populations. Compared with reverse-transcriptase polymerase chain reaction, the test sensitivity was low for all populations (median, 27%; range, 19%-32%), whereas the specificity was high (median, 97%; range, 96%-99.6%).

摘要

使用QuickVue甲型和乙型流感检测(Quidel)对来自3个不同人群的流感样疾病患者的鼻拭子标本进行检测。与逆转录聚合酶链反应相比,该检测对所有人群的敏感性较低(中位数为27%;范围为19%-32%),而特异性较高(中位数为97%;范围为96%-99.6%)。