Department of Medicine and Aging Sciences, Section of Ophthalmology, University G. d'Annunzio, Chieti-Pescara, Chieti, Italy.
Eye (Lond). 2010 Feb;24(2):251-8. doi: 10.1038/eye.2009.76. Epub 2009 Apr 24.
To compare a time-domain (Stratus) and a spectral-domain (Spectralis) optical coherence tomography (OCT) device in assessing foveal thickness in healthy subjects.
In this observational study 40 healthy subjects (40 eyes) underwent Stratus OCT and Spectralis OCT measurements of foveal thickness using three consecutive horizontal and vertical B-scan. Paired samples t-test was used to compare means between Stratus and Spectralis OCT measurements. Coefficient of variation (CoV) was used to compare dispersion in datasets. Pearson's correlation coefficient was used to quantify linear relation between Spectralis and Stratus OCT measurements. To assess agreement between Spectralis and Stratus OCT foveal thickness measurements, the Bland and Altman plots were used.
Sample age ranged from 19 to 49 years (mean 33.25, standard deviation (SD) +/-4.22). The Spectralis OCT foveal thickness measurements resulted significantly higher than those obtained with Stratus OCT (227.64+/-11.74 vs 144.36+/-12.25 microm, and 227.63+/-11.43 vs 144.92+/-12.34 microm, for horizontal and vertical foveal thickness, respectively) (P<0.05). Coefficient of variations were 5.16 and 5.02% using Spectralis OCT, and 8.49 and 8.51% using Stratus OCT. Mean Spectralis/Stratus ratio was 1.58 for both horizontal and vertical measurements. A linear relation between the two technologies was found (r(horiz)=0.899 and r(vert)=0.869) (P<0.001).
A good correlation between Stratus and Spectralis OCT foveal measurements was found, independently of retinal thickness. This preliminary study suggests the existence of a conversion factor between Stratus and Spectralis OCT when measuring healthy foveal thickness.
比较时域(Stratus)和频域(Spectralis)光学相干断层扫描(OCT)设备在评估健康受试者的中心凹厚度中的作用。
在这项观察性研究中,40 名健康受试者(40 只眼)接受了 Stratus OCT 和 Spectralis OCT 中心凹厚度的三次连续水平和垂直 B 扫描测量。采用配对样本 t 检验比较 Stratus 和 Spectralis OCT 测量值之间的平均值。变异系数(CoV)用于比较数据集的离散度。Pearson 相关系数用于量化 Spectralis 和 Stratus OCT 测量值之间的线性关系。为了评估 Spectralis 和 Stratus OCT 中心凹厚度测量值之间的一致性,使用 Bland 和 Altman 图。
样本年龄范围为 19 至 49 岁(平均 33.25,标准差(SD)+/-4.22)。Spectralis OCT 中心凹厚度测量值明显高于 Stratus OCT 测量值(水平为 227.64+/-11.74 与 144.36+/-12.25 微米,垂直为 227.63+/-11.43 与 144.92+/-12.34 微米)(P<0.05)。使用 Spectralis OCT 的变异系数分别为 5.16%和 5.02%,使用 Stratus OCT 的变异系数分别为 8.49%和 8.51%。两种技术的平均 Spectralis/Stratus 比值均为 1.58,水平和垂直测量均为 1.58。两种技术之间存在线性关系(r(水平)=0.899 和 r(垂直)=0.869)(P<0.001)。
Stratus 和 Spectralis OCT 中心凹测量值之间存在良好的相关性,与视网膜厚度无关。这项初步研究表明,在测量健康中心凹厚度时,Stratus 和 Spectralis OCT 之间存在转换因子。