Retina. 2013 Jul-Aug;33(7):1448-55. doi: 10.1097/IAE.0b013e3182807683.
To evaluate the accuracy and reproducibility of retinal thickness measurements in healthy and epiretinal membranes (ERM) eyes by Cirrus and Spectralis Fourier-domain optical coherence tomography devices.
Eighty-seven ERM and 122 healthy subjects underwent 3 macular scans using both optical coherence tomography instruments. Mean thickness measurements in the nine Early Treatment Diabetic Retinopathy Study areas were compared, evaluating the repeatability and the relationship between devices.
Macular thickness increase was detected in ERM eyes for all optical coherence tomography parameters (P < 0.001). Mean foveal thickness was 423.5 ± 81.4 and 438.0 ± 54.2 μm for ERM eyes and 267.1 ± 20.2 and 277.5 ± 18.9 μm for healthy eyes using Cirrus and Spectralis, respectively. Macular average thickness in ERM eyes as determined by both optical coherence tomography was correlated (r = 0.812; P < 0.001) but significantly different (P = 0.044). In ERM eyes, measurements showed a mean of the coefficients of variation of 2.95%, 2.2%, and 1.01% using Cirrus, Spectralis, and Spectralis progression feature, respectively. Intraclass correlation coefficients were higher than 0.919 in all cases.
Reproducibility of both Cirrus and Spectralis optical coherence tomography was high in healthy and ERM eyes. However, considerable differences were found between macular thickness measurements obtained by both devices despite the high correlation between them.
评估 Cirrus 和 Spectralis 频域光学相干断层扫描设备在健康眼和视网膜前膜(ERM)眼中测量视网膜厚度的准确性和可重复性。
87 例 ERM 患者和 122 例健康受试者分别使用两种光学相干断层扫描仪进行 3 次黄斑扫描。比较了 9 个早期糖尿病视网膜病变研究区域的平均厚度测量值,评估了重复性和设备之间的关系。
ERM 眼中所有光学相干断层扫描参数均检测到黄斑厚度增加(P < 0.001)。Cirrus 和 Spectralis 分别测量 ERM 眼的中心凹平均厚度为 423.5 ± 81.4 μm 和 438.0 ± 54.2 μm,健康眼为 267.1 ± 20.2 μm 和 277.5 ± 18.9 μm。两种光学相干断层扫描仪测量的 ERM 眼黄斑平均厚度呈高度相关(r = 0.812;P < 0.001),但差异有统计学意义(P = 0.044)。在 ERM 眼中,Cirrus、Spectralis 和 Spectralis 进展特征分别测量的变异系数平均值为 2.95%、2.2%和 1.01%。所有情况下的组内相关系数均高于 0.919。
Cirrus 和 Spectralis 光学相干断层扫描仪在健康眼和 ERM 眼中的重复性均较高。然而,尽管两种设备之间存在高度相关性,但仍发现黄斑厚度测量值之间存在显著差异。